Page 10
For technical questions, please call 1-888-866-5797.
Item 64019
S
AFETY
Op
ERA
TION
M
AINTENANCE
S
ETU
p
Measurement Operation
Note:
Remove plugs from
ends of Test Leads (included)
before connecting to Meter.
Note:
Test Lead probes have removable
covers for overvoltage protection. With
covers in place, Test Leads are rated for
use with CAT IV circuits. Exposed probes
are rated for use with CAT II circuits.
AC/DC Voltage Measurement
Measure AC conductors carrying
up to 1000 VAC, 40-400 Hz.
Measure DC conductors
carrying up to 1000 VDC.
WARNING!
Use caution when working
near voltages above 30 VAC rms,
42 VAC peak, or 60 VDC. Voltages this
high present a risk of electric shock.
1. Plug black test lead into
COM
Jack.
Plug red test lead into
Input
Jack.
2. Turn Rotary Switch to the
position.
3. Press
FUNC
button to choose
between AC and DC voltage.
4. Carefully touch exposed
conductors with tips of probes.
5. Read measured voltage on the
Display. For AC voltage, frequency
is shown at the top right.
• Use the
Lo-Z
switch position on
Meter when readings are suspect
(ghost or stray voltages may
be present), or when testing for
the presence of voltage. Follow
instructions for
“Low Impedance
Measurement” on page 14.
6. When testing is complete, turn
Rotary Switch to
OFF
, remove
Test Leads and store with Meter.
Frequency/Duty Cycle Measurement
Measure frequency up to 66MHz.
1. Plug black test lead into
COM
Jack.
Plug red test lead into
Input
Jack.
2. Turn the Rotary Switch to
the
Hz%
position.
3. Connect the Test Leads across
the circuit to be measured.
4. Read measured frequency in
the center of the Display. Duty
cycle is shown at the top right.
5. When testing is complete, turn
Rotary Switch to
OFF,
remove
Test Leads
and store with Meter.