Pump P-910
Instructions 56-3000-47 Edition AD

8 Measure the volume collected in

the measuring cylinder. Enter this value in

Reference field

. Click on 

Set ref value

then

click on 

Save

.

9 The pump is now calibrated.

Storage

Overnight: The pump can be left filled with a
buffer.

Weekend and Long time storage: Flush the
pump with water and then fill it with 20%
Ethanol.

Note: If the pump tubing is left in the pump
for a long period it will become deformed
and will not function correctly. 
For long time storage remove the pump
tubing from the pump to avoid deformation
of the tubing.

4 Maintenance

CAUTION! Only spare parts approved or

supplied by Amersham Pharmacia Biotech
may be used for maintaining and servicing
Pump P–910.

Period

Action

If tubing breaks

Perform cleaning procedure
detailed below.

Every 200 hrs

Replace pump tubing. See 
section 2.

Cleaning-in-place

Pump a cleaning or sanitizing agent through
the pump. The standard recommendation is
to pump 1 M NaOH for 30 minutes and then
wash out with buffer.

WARNING! NaOH is injurious to health.
Avoid spillage.

Cleaning procedure

1 Open the door and remove the tubing.

2 Dismantle the roller cage by undoing

the Allen screw and lifting the cage
straight out. Take care not to lose the two
metal washers.

3 Slide the pressure plate forward and

remove it from the pump.

4 Clean the roller cage and the roller cage

housing with soapy water.

5 If the roller cage O-ring has come into

contact with the eluent, replace the O-ring.

6 Re-assemble the pump by reversing

the dismantling procedure.

5 Trouble shooting

If the suggested actions do not correct the
fault, call Amersham Pharmacia Biotech.

Fault

Action

Leakage

1 Check the top and bottom pump

tubing connections.

2 Check if there is damage to the

pump tubing. Replace if necessary.

Not running

Check the UniNet connections.

56-3000-47_AD.qxd  1998-10-21 14:17  Sida 3 (1,1)

Содержание P-910

Страница 1: ...complete system Unpacking Unpack the instrument and check the items against the packing list Inspect the items for obvious damage which may have occurred during transportation Installing the pump At...

Страница 2: ...a method or manually via the functions available in UNICORN Setting the flow rate 1 Select menu System Control Manual Pump in UNICORN 2 Select the instruction SampleFlow in the list 3 Enter the requir...

Страница 3: ...ee section 2 Cleaning in place Pump a cleaning or sanitizing agent through the pump The standard recommendation is to pump 1 M NaOH for 30 minutes and then wash out with buffer WARNING NaOH is injurio...

Страница 4: ...the roller bearings As the cage rotates the rollers rotate in the opposite direction driven by drive wheels which engage the chassis ring The roller diameter and the gear ratio ensure that the roller...

Страница 5: ...declaration of conformity is valid for the instrument when it is used in laboratory locations used in the same state as it was delivered from Amersham Pharmacia Biotech except for alterations describe...

Страница 6: ...6 Pump P 910 Instructions 56 3000 47 Edition AD 56 3000 47_AD qxd 1998 10 21 14 17 Sida 6 1 1...

Страница 7: ...7 Pump P 910 Instructions 56 3000 47 Edition AD 56 3000 47_AD qxd 1998 10 21 14 17 Sida 7 1 1...

Страница 8: ...otech AB SE 751 84 Uppsala Sweden Trademarks KTA and UNICORN are trademarks of Amersham Pharmacia Biotech Limited or its subsidiaries Amersham is a trademark of Nycomed Amersham plc Pharmacia and Drop...

Отзывы: