2

Pump P-910

Instructions 56-3000-47 Edition AD

Inserting the tubing

1 Open the door.

2 Push back the

pressure plate.

3 Insert the tubing.

Make sure 
the nuts of 
the connectors
fit correctly
into 
the housing.

4 Close the door. The correct tubing

pressure is set automatically.

Connecting the pump

Pump P-910 is connected to the system with
1/16” diameter PTFE or PEEK tubing, via the
two connectors on the pump tubing. 
Connect the outlet
to the upper
connector,
marked 1, and 
the inlet to the lower connector, marked 2, as
follows:

1 Cut the tube cleanly at a 45°angle.

2 Remove the connector nut (1) and slip it

over the tubing.

3 Remove the sealing plug (2).

4 Push the tubing through the sealing plug

until it protrudes about 1 cm.

5 Seat the tubing in the nipple (3). Slide

the sealing plug along the tubing onto
the nipple.

6 Fingertighten the connector nut. 

3 Operation

Pump P-910 is controlled from a PC running
UNICORN version 2.10 or higher. It cannot
be used as a stand-alone instrument. Control
of the pump can be achieved automatically
from a method, or manually via the functions
available in UNICORN.

Setting the flow rate

1 Select menu 

System Control:Manual:Pump 

in

UNICORN.

2 Select the instruction 

SampleFlow 

in

the list.

3 Enter the required flow rate.

4 Click on the 

Execute

button. The pump

starts.

5 To stop the pump, set the flow rate to 0.

The pump can be run without liquid, but it
should be avoided due to mechanical wear.

Calibrating the pump

The pump must be calibrated every second
day when in daily use or whenever the
running conditions are changed, e.g. viscosity
of sample or buffer, temperature, back-
pressure, etc. If the pump is not used
frequently it should be calibrated each time.

1 Select menu 

System Control: System

Calibrate 

in UNICORN.

2 Select 

SamplePump.

3 Place the inlet tubing in a bottle of water.

4 Start the pump by selecting 

Manual:Pump

and 

Sampleflow

. Enter a suitable flow rate,

e.g. 2 ml/min. Click on 

Execute.

5 Place the outlet tubing from the pump in a

10 ml measuring cylinder, at the same
time pressing 

Start collect

in the 

Calibration

box.

6 Let the pump run for 2 min.

7 Press 

Stop collect

, at the same time

removing the outlet tubing from
the measuring cylinder.

1

2

3

1

2

Pressure plate

Roller cage

56-3000-47_AD.qxd  1998-10-21 14:17  Sida 2 (1,1)

Содержание P-910

Страница 1: ...complete system Unpacking Unpack the instrument and check the items against the packing list Inspect the items for obvious damage which may have occurred during transportation Installing the pump At...

Страница 2: ...a method or manually via the functions available in UNICORN Setting the flow rate 1 Select menu System Control Manual Pump in UNICORN 2 Select the instruction SampleFlow in the list 3 Enter the requir...

Страница 3: ...ee section 2 Cleaning in place Pump a cleaning or sanitizing agent through the pump The standard recommendation is to pump 1 M NaOH for 30 minutes and then wash out with buffer WARNING NaOH is injurio...

Страница 4: ...the roller bearings As the cage rotates the rollers rotate in the opposite direction driven by drive wheels which engage the chassis ring The roller diameter and the gear ratio ensure that the roller...

Страница 5: ...declaration of conformity is valid for the instrument when it is used in laboratory locations used in the same state as it was delivered from Amersham Pharmacia Biotech except for alterations describe...

Страница 6: ...6 Pump P 910 Instructions 56 3000 47 Edition AD 56 3000 47_AD qxd 1998 10 21 14 17 Sida 6 1 1...

Страница 7: ...7 Pump P 910 Instructions 56 3000 47 Edition AD 56 3000 47_AD qxd 1998 10 21 14 17 Sida 7 1 1...

Страница 8: ...otech AB SE 751 84 Uppsala Sweden Trademarks KTA and UNICORN are trademarks of Amersham Pharmacia Biotech Limited or its subsidiaries Amersham is a trademark of Nycomed Amersham plc Pharmacia and Drop...

Отзывы: