17
limit the voltage. This means running at constant voltage for the last phase (phase III) to prevent local
overheating.
For detailed information about parameter values, always follow the gel supplier´s recommendations.
5.7 Short instructions
This section summarizes the main programming points covered earlier in this chapter. Use it as a check
list once you are familiar with the detailed programming, editing and running instructions. Refer also to
the separate schematic operating guide included with the power supply. We recommend you keep this
schematic guide close by the instrument.
1. Turn mains power ON. The display should have the program number flashing.
2. Press SET ENTER for the actual program number, or use
Ô
/
Õ
to select the program number desired.
Press SET ENTER.
3. The display now asks if you wish to perform a step or gradient program. Select
with SET ENTER,
or with
Ô
/
Õ
followed by SET ENTER.
4. The display will blink for phase 1. Press SET ENTER to confirm.
5. For each of the parameters Voltage, Current and Power, press
Ô
/
Õ
until the value desired is reached.
If gradient mode has been selected, the voltage endpoint for the phase should be chosen. Confirm by
pressing SET ENTER after each parameter, and the display will automatically move to the next
parameter.
6. Choose between manual break or automatic break. For automatic break, choose breakpoint unit in
either time (h), volthours (Vh) or milli-amperehours (mAh). Select the correct unit or, for manual
break, choose
“OFF” with
Ô
/
Õ
. Confirm with SET ENTER.
7. If hours, volthours or milliamperehours were chosen, set the value to be used for the breakpoint with
Ô
/
Õ
and SET ENTER.
8. If you only want one phase, press SET ENTER for YES to confirm that this is the last phase. If more
phases are required, select NO with ‘/’ and confirm with SET ENTER. To add more phases, repeat
steps 5–8, until the desired number of phases have been programmed. Up to 9 phases can be
programmed.
I
II
III
IV
Power
Voltage
Current
Preset time
Alarm
Fig. 9. Changes in power, voltage and current during an IEF run.
EPS 3501 XL - User Manual
18-1130-22 AA EPS 3500XL inlaga 99-02-10 13.09 Sidan 17
Содержание EPS 3501 XL
Страница 36: ...14 EPS 3501 XL Manuel d Utilisation 18 1130 22 AA EPS 3500XL inlaga 99 02 10 13 10 Sidan 34 ...
Страница 50: ...14 EPS 3501 XL Bedienungsanleitung 18 1130 22 AA EPS 3500XL inlaga 99 02 10 13 10 Sidan 48 ...
Страница 64: ...14 EPS 3501 XL Manual del usuario 18 1130 22 AA EPS 3500XL inlaga 99 02 10 13 10 Sidan 62 ...
Страница 78: ...14 18 1130 22 AA EPS 3500XL inlaga 99 02 10 13 10 Sidan 76 ...
Страница 79: ...18 1130 22 AA EPS 3500XL omslag 99 02 10 13 17 Sidan 3 ...