44
098-0010 Rev C
Output High
Pressure
Warning
Delivery Pressure Switch (PS1)
registers value higher than set-
point on pressure switch
X12 off
1) Permeate valve relief set too high
2) Check pressure switch set-point by
comparing gauge reading when warning
occurs
1) Adjust valve using Section 9.1
2) Adjust set-point
High Water
Temperature
Warning
Loop Over Temp Warning Set-
point Exceeded
1) Check Loop Over Temp Warning set-
point
2) Check loop temperature
1) Adjust warning set-point
Output Quality
Warning
Output Quality Warning Set-point
Exceeded
1) Check Output Quality Warning set-point
2) Check Output Conductivity
3) Check conductivity with myron-l meter
1) Check line cell
2) Perform chemical clean
3) Replace membrane
Output Quality
out of Range
Warning
Output Conductivity Sensor
transmitting the maximum
detectable value for 30 seconds
1) Check sensor and wiring
2) Check conductivity board
3) Check cables between board and PLC
1) Check line cell
2) Perform chemical clean
Output Line
Cell Error
Warning
Output Conductivity Sensor
transmitting over the maximum
detectable value for 30 seconds
1) Check sensor and wiring
2) Check conductivity board
3) Check cables between board and PLC
1) Check line cell
2) Perform chemical clean
Tank Quality
Warning
Tank Quality Warning Set-point
Exceeded
1) Check Tank Quality Warning set-point
2) Check conductivity with Myron-l meter
3) Check water softener and any
pretreatment tanks
4) Check tank sensor and wiring
1) Check line cell
Tank Quality
out of Range
Warning
Tank Conductivity Sensor
transmitting the maximum
detectable value for 30 seconds
1) Check sensor and wiring
2) Check conductivity board
3) Check cables between board and PLC
1) Check line cell
Tank Line Cell
Error Warning
Tank Conductivity Sensor
transmitting over the maximum
detectable value for 30 seconds
1) Check sensor and wiring
2) Check conductivity board
3) Check cables between board and PLC
1) Check line cell
Loop Supply
Sampling Error
Warning (MDP
only)
During operation, the Stage 1
Conductivity Sensor is detecting
value greater than Loop Quality
Warning Set-point for 30 seconds
1) Check the Stage 1 conductivity
1) Perform membrane chem clean
2) Replace membrane
Содержание MediQA
Страница 2: ......
Страница 4: ...2 098 0010 Rev C 1 COMPONENT IDENTIFICATION 1 1 MSP Components ...
Страница 5: ...3 098 0010 Rev C ...
Страница 6: ...4 098 0010 Rev C ...
Страница 7: ...5 098 0010 Rev C 1 2 MDP Components ...
Страница 8: ...6 098 0010 Rev C ...
Страница 9: ...7 098 0010 Rev C ...
Страница 11: ...9 098 0010 Rev C 2 2 Flow Schematics 2 2 1 MSP Schematics ...
Страница 12: ...10 098 0010 Rev C ...
Страница 13: ...11 098 0010 Rev C ...
Страница 14: ...12 098 0010 Rev C ...
Страница 15: ...13 098 0010 Rev C ...
Страница 16: ...14 098 0010 Rev C ...
Страница 17: ...15 098 0010 Rev C ...
Страница 18: ...16 098 0010 Rev C ...
Страница 19: ...17 098 0010 Rev C ...
Страница 20: ...18 098 0010 Rev C ...
Страница 21: ...19 098 0010 Rev C ...
Страница 22: ...20 098 0010 Rev C ...
Страница 23: ...21 098 0010 Rev C ...
Страница 24: ...22 098 0010 Rev C ...
Страница 25: ...23 098 0010 Rev C 2 2 2 MDP Schematics ...
Страница 26: ...24 098 0010 Rev C ...
Страница 27: ...25 098 0010 Rev C ...
Страница 28: ...26 098 0010 Rev C ...
Страница 29: ...27 098 0010 Rev C ...
Страница 30: ...28 098 0010 Rev C ...
Страница 31: ...29 098 0010 Rev C ...
Страница 32: ...30 098 0010 Rev C ...
Страница 33: ...31 098 0010 Rev C ...
Страница 34: ...32 098 0010 Rev C ...
Страница 35: ...33 098 0010 Rev C ...
Страница 36: ...34 098 0010 Rev C ...
Страница 37: ...35 098 0010 Rev C ...
Страница 51: ...49 098 0010 Rev C 8 COMPONENT FAILURES ...