14
098-0007 Rev B
4.4
Environment
The device should be used indoors on a firm, flat level surface in a clean and dry environment. The
device should be operated in a well ventilated area. Take care when making water and power
connections to the rear of the device as to not create tripping hazards. Refer to Section 2.3 for safety
considerations when installing the Centurion Plus Heat disinfection device.
4.5
Checking Water Services
The Centurion Plus Heat disinfection device will use a portable RO system as the feed water supply. The
RO system must be capable of being connected to an 8mm/5/16” push-fit connection. It is recommended
that the RO have a recirculation line back from the Hemodialysis machine.
The Centurion Plus Heat disinfection system will accept the reject water from an RO using an 8mm/5/16”
push-fit connection. The reject water will pass through the Centurion Plus Heat disinfection device to a
suitable drain. The Reject water will require a suitable drain. See Section 5.2.2 for drain requirements.
WARNING:
The drain must be capable of handling water that could be up to 203 °F.
4.5.1
Water connection detail
NOTE:
Water connections should be placed such that they are not facing the operator during disinfection
cycle.
CENTURION PLUS
Содержание 000-HS120
Страница 2: ......
Страница 24: ...22 098 0007 Rev B Right Side...
Страница 25: ...23 098 0007 Rev B Left Side Top with top panel and controller removed...
Страница 27: ...25 098 0007 Rev B 6 3 Flow Schematic 6 3 1 Standby Mode...
Страница 28: ...26 098 0007 Rev B 6 3 2 Preheat Mode...
Страница 29: ...27 098 0007 Rev B 6 3 3 Dispensing Mode...
Страница 30: ...28 098 0007 Rev B 6 3 4 Drain Mode...
Страница 68: ...66 098 0007 Rev B 13 SPARE PART LISTING 13 1 Exploded View...
Страница 69: ...67 098 0007 Rev B...
Страница 70: ...68 098 0007 Rev B...
Страница 71: ...69 098 0007 Rev B...
Страница 72: ...70 098 0007 Rev B...
Страница 73: ...71 098 0007 Rev B...
Страница 74: ...72 098 0007 Rev B...