AMI
AMI MODEL 4Q05100PS
4-QUADRANT POWER SUPPLY
INSTALLATION, OPERATION, AND
MAINTENANCE INSTRUCTIONS
I.
INTRODUCTION
The AMI Model 4Q05100PS is designed to provide a compact, 4-quadrant capability for
AMI power supply systems. The Model 4Q05100PS is designed to operate with the AMI
Model 420 Power Supply Programmer.
II. SPECIFICATIONS
________________________________________________________________________________
AC Input Voltage: . . . . . . . . . . . . . . . 100-120 VAC or 200-240 VAC ±10%
1
AC Input Frequency: . . . . . . . . . . . . . 47 to 63 Hz
DC Output Current/Compliance: . . . ±100 A @ ±4.0 V, ±80 A @ ±4.7 V, ±60 A @ ±5.0 V
Power-Absorbing Mode
2
Rating: . . . . 325 VA max. @ 25°C, linearly derated to
295 VA max. @ 40°C
Output Overvoltage Protection: . . . . ±6 VDC (nominal)
Ambient Temperature Rating . . . . . . 0 to 40°C
352*5$0,1
Control Signal:
. . . . . . −
10 to +10 VDC
3
Output Voltage Noise: . . . . . . . . . . . . 0.6 mV RMS into 20 MHz BW (typical)
10 mV p-p into 20 MHz BW (typical)
75 mV p-p into 20 MHz BW (maximum)
Torque Limit on Current Terminals: 50 in-lbs.
________________________________________________________________________________
1
Input voltage range is selectable via an internal switch.
2
The supply is in absorbing mode when the polarity of the output voltage is opposite the
polarity of the current flow.
3
Gain of 0.5 is applied to the voltage output of the unit.