Demarcation of Lens

 

ƒ

 

For demarcation of a spherical lens, rotate the diopter measuring handwheel to focus the lines of 
the (3) marker reticule clearly.  Lightly move the lens to find the right optical center.  Make three-
point printing marks with the printer.   

ƒ

 

For demarcation of the axis of a sphero-cylindrical lens, rotate the astigmatism axis measuring 
handwheel to the required angle.  Then rotate the diopter measuring handwheel while turning the 
lens to focus the cylindrical marking lines clearly.  Lightly move the lens to find the right optical 
center.  Make three-point printing marks with the printer. 

ƒ

 

For demarcation prismatic lens base, rotate the astigmatism axis measuring handwheel to the 
required angle.  Then rotate the diopter measuring handwheel to focus the lines of the (3) marker 
reticule clearly.  Lightly turn the prism lens so that the middle of the three green marking lines 
passes through the center of the (8) eyepiece reticule.  Make printing point marks by the use of 
the printer. 

 
 

MAINTENANCE

 

ƒ

 

Do not dismantle the lensometer, as the accuracy will be compromised.  

ƒ

 

Store the lensometer in a dry, cool, indoor place in order to assure precise working condition. 

ƒ

 

Prevent the instrument from severe vibration or impact in order to avoid damage or loosening parts. 

ƒ

 

When changing the bulb, make sure the bulb is cool before touching. 

ƒ

 

Always keep the instrument clean.  Do not touch the surfaces of the optical parts.  Dust shall be cleaned 
with an absorbent cotton-tipped applicator dipped with a mixture of ethanol and ether. 

ƒ

 

After each use, clean the instrument and cover it with a dust shield. 

 

 

COMMON TROUBLE SHOOTING

 

Problem Reason 

Solution 

 

Bulb will not light up. 

 

 

Mains are not connected. 

 

Connect with the mains. 

 

Cannot see eyepiece reticule when 

adjusting the eyepiece readings. 

 

 

Observer has too much 

uncorrected refractive error. 

 

Wear a refractive correction 

when using the lensometer. 

 

Green marking image cannot be 

seen. 

 

 

Bulb cannot be lighted or 

Dust shield is still on. 

 

 

Connect with mains or  

Take off dust shield. 

 

Green marking image 

cannot be seen clearly. 

 

 

Dust accumulates  

on optical lens. 

 

Clean it using cotton  

dipped in cleaning solution. 

 

 

Deviation of the center 

of the green marking image. 

 

 

Lock screws are loose. 

 

Remove objective lens shield, 

regulate three screws, lock them. 

 

 
 

Translocation of the printing unit. 

 

 

Loosening of limit 

screw or loosening of 

connecting screws. 

 

Regulate limit screw & lock 

or regulate connecting 

screws & lock. 

 

Содержание EQ-6001

Страница 1: ...1 800 255 6161 Fax 1 800 397 0013 www amconlabs com AMCON LENSOMETER USER S MANUAL EQ 6001...

Страница 2: ...it 5 Objective Lens Bearing Seat 6 Diopter Measuring Handwheels 6 Astigmatism Axis Measuring Handwheel 6 Illuminating Bulb Chamber 7 Instrument s Inclination Regulating Handle 7 Operation Instruction...

Страница 3: ...ll Dimensions of Device 108 25 in X 51 25 in X 179 in Weight 12 5 lbs Lighting Lamp 220V 110V 15W PRINCIPLE OF OPERATION This instrument consists of two coaxial optical systems Light from the 1 light...

Страница 4: ...spiral focusing unit whose range of focusing is 5D so as to suit various refractive errors Fig 2 Eyepiece Reticule and Marker Reticule Fig 3 Fig 4 Reticule in the field of view of eyepiece is shown as...

Страница 5: ...val of 0 125D Those between 5D and 25D are at an interval of 0 25D Prism Compensation Device Fig 6 The 6 prism compensation device is needed when measuring a lens diopter value above 5 There are two l...

Страница 6: ...hing handle forward to withdraw the lens pushing board During measurement first press the lens with the lens pressing unit and slightly move the lens so that it aims right at the center Then fix it wi...

Страница 7: ...Handwheels Fig 11 The diopter measuring handwheels are located on either side of the lensometer They can be rotated smoothly and steadily to adjust the focus until the observer can clearly see the ind...

Страница 8: ...asurement Place the instrument on the working table and adjust the inclination so that the operator is able to make the measurement comfortably Connect with the mains Before measurement visibility of...

Страница 9: ...trument is used to measure such kind of lens two green lines and three green lines of 3 marker reticule will not get clear simultaneously The measurement of this lens takes two steps First step Rotate...

Страница 10: ...cal center at the center of the optical axis Rotate the diopter measuring handwheel to focus the lines of the 3 marker reticule clearly The marking lines can be seen to deviate from the center of 8 ey...

Страница 11: ...r place in order to assure precise working condition Prevent the instrument from severe vibration or impact in order to avoid damage or loosening parts When changing the bulb make sure the bulb is coo...

Отзывы: