3. Specifications
The Ambu Res-Cue Pump is in conformity with the
following standard: EN ISO 10079-2:1996 E.
The Ambu Res-Cue Pump is in conformity with Council
Directive 93/42/EEC concerning Medical Devices
Vacuum (max):
-450 mmHg
Peak free airflow:
> 20 l/min
Disposable container volume:
300 ml
External diameter of cap (suction port):
17 mm
External diameter of catheter connection:
13 mm and
6 mm
Stroke Adjustment:
Turn stroke adjustment knob 90°
counter-clockwise to generate 50% vacuum (of original
pressure).
Operating environmental temperature:
-20 °C to + 50 °C
(-4 °F to 122 °F)
Storage environmental temperature:
-40 °C to + 60 °C
(-40 °F to 140 °F)
Dimensions:
185 x 70 x 168 mm
Weight:
260 g
4. Description
Ambu Res-Cue Pump is a light portable suction unit that
can be powered by one hand and allows the other hand
free for other important duties. This suction unit was
designed to provide simple operation and maintenance.
The adjustable stroke knob will provide different suction
pressures.
5. Operating instructions
Select appropriate size tubing (according to the size of the
patient) and attach it to the adapter (if not attached
already). Attach the adapter with tubing to the cap of the
container. The container should be attached to the pump.
Caution:
It is very important that the container cap is
attached all the way to the pump body.
The Res-Cue Pump is operated by a simple squeezing
action of the trigger. The adjustment stroke knob will
enable it to produce various suction volumes (see below).
Pull or squeeze the trigger toward the pump handle. This
squeezing mechanism will generate the vacuum in the
pump and suck the fluids into the
container. Once suction is completed, remove tubing slow-
ly from the patient. Remove filled container by unsnapping
the container (and tubing) from the pump. Place container
along with the tubing in the poly bag. Discard disposable
tubing and container appropriately.
English
5