7.54
Second analog output operation mode
Display:
rEC2
Range:
rAtE
or
totL
Default Value:
rAtE
Seen only when the
SEt OPER
function is set to
both
. Sets the operation mode for the second
optional analog retransmission. The output can be set to retransmit the total (
totL
) or rate
(
rAtE
).
7.55
Bargraph display operation mode
Display:
bAR
Range:
rAtE
or
totL
Default Value:
rAtE
Seen only when
SEt OPER
is set to
both
. Allows the choice of rate or total to be displayed on
the bargraph in bargraph model displays.
7.56
Digital output operation mode
Display:
dGOP
Range:
rAtE
or
totL
Default Value:
rAtE
Seen only when
SEt OPER
is set to
both
. Allows the choice of rate or total to be output on
the optional digital output.
7.57
Low overrange visual warning limit value
Display:
Lo dISP
Range:
Any display value or
OFF
Default Value:
OFF
Low overrange limit value - the display can be set to show an overrange message if the display
value falls below the
Lo dISP
setting. For example if
Lo dISP
is set to
50
then once the
display reading falls below
50
the message
-or-
will flash on and off or the display value will
flash on and off instead of the normal display units (see
dISP
function
). This message can
be used to alert operators to the presence of an input which is below the low limit. If this function
is not required it should be set to
OFF
by pressing the
^
and
v
buttons simultaneously at this
function.
Note: this function operates on frequency/rate display only.
48 of
PM4TRMAN-2.7-0