5.12
Analog retransmission outputs
Section:
RO 1
to
RO 2
Display:
OutP
Range:
4-20
,
0-1.0
,
0-10
Default Value:
4-20
Default Access Level
4
Function number
4140
to
4141
One or two analog outputs are optionally available in either 12 or 16 bit versions. The 12 bit version output is fixed
at 4-20mA. With the 16 bit version the user can select 4-20mA, 0-1VDC or 0-10VDC output at this function.
5.13
Analog retransmission input channel
Section:
RO 1
to
RO 2
Display:
InPt
Range:
tMR
Default Value:
tMR
Default Access Level
4
Function number
43E0
to
43E1
This function allows selection of which channel the selected analog output is to follow. In this instrument the only
choice is timer operation
tMR
5.14
Analog retransmission low display value
Section:
RO 1
to
RO 2
Display:
Lo
Range:
Any display value
Default Value:
0.0
Default Access Level
4
Function number
4120
to
4121
This function can be used to set the analog retransmission signal output low value in displayed engineering units.
For example to set analog output 1 to retransmit 4mA (or 0V if available) for a display value of zero set
RO1 Lo
to
0
.
5.15
Analog retransmission high display value
Section:
RO 1
to
RO 2
Display:
High
Range:
Any display value
Default Value:
100.0
Default Access Level
4
Function number
4130
to
4131
This function can be used to set the analog retransmission signal output high value in displayed engineering units.
For example to set analog output 1 to retransmit 20mA (or 1V or 10V if available) for a display value of 200 set
RO1 Lo
to
200
.