![Amada TC-W100A Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/amada/tc-w100a/tc-w100a_operation-manual_2915854025.webp)
TC-W100A - THERMOCOUPLE WELDER
990-070
4-1
CHAPTER 4
USER MAINTENANCE
Section I. Handpiece Maintenance
Wire Clamp Jaws
With use, the Upper Wire Clamp Jaw will accumulate wire insulation and metal particles inside the Guide
Tube. The buildup rate depends on the number of thermocouple welds made, the thermocouple insulation
material, and the thermocouple wire material. The larger size wires will accumulate particles at a faster rate.
To clean both Wire Clamps, remove them from the Handpiece and place them in an ultrasonic cleaner. Add
a cleaning solvent such as the Cole-Parmer Micro-90 Cleaning Solution.
WARNING:
Do not use a wire to clean the upper Wire Clamp guide tube. Doing so may cause internal
damage to the guide tube.
Handpiece Cable Replacement
The Handpiece Cable is very flexible and should last many years with careful use. Should the insulation
become cracked at either the Handpiece or Power Supply front attachments, return the unit to AMADA
WELD TECH. See
Chapter 5, Section III, Repair Service
.
Содержание TC-W100A
Страница 1: ...THERMOCOUPLE WELDER TC W100A OPERATION MANUAL 990 070 Rev H...
Страница 12: ......
Страница 18: ......
Страница 30: ...AMADA WELD TECH INC 1820 South Myrtle Ave Monrovia CA 91016 U S A TEL 1 626 303 5676 FAX 1 626 358 8048...