background image

-16-

E.

Press the setup key until the display shown below appears.  The sensor voltage value for
the minimum open position can now be entered.  Recall the value read and recorded in

 

section 3.5, step B.

 

This value

 

plus 

5

0 millivolts should be used for the MINPOS

 

value.

 

The +

50

 millivolt offset voltage allows for a 0.00% output reading from the

 

valve to any

 

user equipment when it is off, in spite of any thermal growth of the valve

 

parts which may

 

occur during normal use.

HARDWARE CONFIG
MINPOS

ÿ

  1.880

 +

pres s to
increase
value

 -

pres s to
decrease
value

The right arrow always points to the adjustable value on the CONFIGURATION
screen.  Adjustments to CONFIG values are saved instantly and retained until
changed again.

F.

Use the UP arrow key to select the screen for the entry of the sensor voltage
representing the maximum open position of the valve or 100% open.  Recall the value
read and recorded in section 3.5 step C.  This value

 

minus

 5

0 millivolts should be used

 

for the MAXPOS value.  The -

50

 millivolt offset voltage ensures that the valve will be able

 

to reach its wide open position regardless of minor system variations.

 

E

pres s to
display
nex t
value

HARDWARE CONFIG
MAXPOS

ÿ

 3.720

 +

pres s to
increase
value

 -

pres s to
decrease
value

NORM

press to
ex it
setup
mode

3.6 After returning from the position sensor calibration screen, follow these instructions for testing

the calibration of the AGV5.

A.

The input simulator should still be set to 4 milliamperes or less and the HOME screen
display should appear as shown below.  The position sensor voltage should be about the
same as read in section 3.5 step B, and the valve position shown in % open should now
read close to the fully closed position of 0.00%. A reading of 0.00 +/- 0.50% would be in
the normal range.

OFF
ACT 1.880 -00.50

Содержание AGV5 Series

Страница 1: ...E AGV5 1 AGV5 1A AGV5 2 AGV5 2A IMPORTANT SAFETY NOTICE PROPER INSTALLATION MAINTENANCE REPAIR AND OPERATION OF THIS EQUIPMENT IS ESSENTIAL THE RECOMMENDED PRACTICES CONTAINED HEREIN SHOULD BE FOLLOWE...

Страница 2: ...SERVICE MANUAL SECTION ITEM PAGE 1 0 VALVE SERVICE AND REPAIR MANUAL OVERVIEW 3 2 0 PARTS IDENTIFICATION 5 3 0 REPLACEMENT OR UPDATE OF PCB ASSEMBLY 13 4 0 REPLACEMENT OF THE PRESSURE TRANSDUCER 17 5...

Страница 3: ...g To facilitate troubleshooting and to establish service schedules a maintenance log should be kept on the valve Calibration Calibration of the AGV5 is performed in a controlled environment before shi...

Страница 4: ...FIG 1A PART IDENTIFICATION WHITE GREEN COVER CONTROL HOUSING ASSEMBLY ACTUATOR ASSEMBLY PORT ASSEMBLY TERMINAL WIRING TOP VIEW 0 GREEN...

Страница 5: ...ION 1 1 881002 KT PCB Field Retrofit Kit 1a 1 872005 PCB Assem bly Logic 1b 1 810114 Bushing Magnet 1c 1 810112 Assembly Magnet Holder 1d 4 801008 W asher Flat 4 1e 4 801009 W asher Lock 4 1f 4 801010...

Страница 6: ...FIG 1 B PART IDENTIFICATION NOTE DRAWING SHOWS CURRENT CONFIGURATION OF VALVE SOME MINOR DIFFERENCES MAY EXIST ON OLDER UNITS PRIOR TO S N 2001...

Страница 7: ...f 1 810036 O ring 2g 1 810037 O ring 2h 1 810027 O ring 2i 1 810025 O ring 2j 1 810031 O ring 2k 1 803020 O ring Lube 3 3 801018 Screw 10 32 4 1 50303076 Valve Seat Primary 5 1 50303086 Valve Seat Sec...

Страница 8: ...FIG 1C PART IDENTIFICATION ACTUATOR HOUSING MAGNET ASSEMBLY POPPET ASSEMBLY...

Страница 9: ...4 801009 W asher Lock 4 1f 4 801010 Screw 4 40 2 1 820020 Valve Repair Kit 2a 1 810033 O ring 2b 1 810038 O ring 2c 1 810035 O ring 2d 2 810023 O ring 2e 1 810024 O ring 2f 1 810036 O ring 2g 1 81003...

Страница 10: ...RT IDENTIFICATION NOTE FOR UPDATE KIT 820022 IT 19 ALL COMPONENTS SHOWN COME PREASSEMBLED AND TESTED 820022 KIT ALSO INCLUDES A SPECIAL MAGNET HOLDER ASSEMBLY WITH A 10 32 THREAD TO FIT VALVES BELOW S...

Страница 11: ...1010 Screw 4 40 2 1 820020 Valve Repair Kit 2a 1 810033 O ring 2b 1 810038 O ring 2c 1 810035 O ring 2d 2 810023 O ring 2e 1 810024 O ring 2f 1 810036 O ring 2g 1 810037 O ring 2h 1 810027 O ring 2i 1...

Страница 12: ...4 20 mA OUTPUT 1 4 20mA SIMULATOR i I A B C D E F G H J K L AGVS DIRECT ACTING GAS VALVE 19 36 VDC POWER 19 36 VDC RETURN _ 19 36 VDC RETURN 4 20 POSITION OUT 4 20 PRESSURE OUT 6 SERIAL RS485 _ SERIAL...

Страница 13: ...to grip the cover C Remove the old PCB assembly 1a by removing the four Phillips headscrews 1f which secure it to the pressure transducer housing base 19 With the PCB assembly 1a loose locate the thr...

Страница 14: ...O rings 2c and 2f in place under the pressure transducer housing assembly 19 Tighten the six Allen screws 26 to 18 20 inch pounds in order to hold the transducer housing assembly 19 to the control ho...

Страница 15: ...op line of the display indicates that the valve is operating in the position control mode and the desired set point from the simulator is for a 100 open condition The XXX XX value is the valve positio...

Страница 16: ...oltage representing the maximum open position of the valve or 100 open Recall the value read and recorded in section 3 5 step C This value minus 50 millivolts should be used for the MAXPOS value The 5...

Страница 17: ...plied in a kit 27 which includes all of the necessary parts NOTE When repairing or updating existing AGV5 units with S N 1400 or lower the obsolete pressure transducer and transducer housing as well a...

Страница 18: ...transducer housing K Attach the wiring to the PCB assembly 1a in the proper orientation 5 0 AGV5 PRESSURE TEST 5 1 After repair and before returning an AGV5 valve to service the valve assembly must be...

Отзывы: