16 | 18
Appendix
Datasheet
Number of measuring channels:
2 (full bridge resistance > 300 ohms)
Supply voltage:
12E - 10 … 18 VDC
24E - 18 … 30 VDC
Electronic protected against voltage reversal
Isolating proof voltage input to output:
200 V
Higher isolated proof voltage on request
Power consumption:
max.5 W
Strain gauge excitation supply:
± 2,5 VDC / ± 5 VDC
Analogue output
0 … 10 V / ±10 V
4 … 20 mA
max. 1 mA (short-period short-circuit proof)
max. 500 Ω
Limit frequency (-3 dB):
1 kHz (10 Hz with low pass filter)
optional up to 30 kHz
Input resistance:
>3 MΩ
Max. input sensitivity:
100 mV/V
at +10 VDC excitation supply
Non-linearity:
±0.05 % FSO
Electrical connection:
Pluggable Screw clamps
Housing:
Plastic enclosure for top hat rail mounting (IP20)
Dimension (B x H x D):
100 x 73,2 x 118,2 mm
Weight:
450 g
Temperature, storage:
-20 °C … +60 °C
Temperature, operating:
0 °C … +50 °C
Order designation
SG-2K-KS…
2 channel strain gauge-amplifier in plastic housing for DIN-top hat rail mounting (IP20)
…-12E-…
Supply voltage: 10 … 18 VDC
…-24E-…
Supply voltage: 18 … 30 VDC
…-010-…
Analogue output: 0 … 10 V
…-B10-…
Analogue output: ± 10 V
…-420-…
Analogue output: 0 … 10 V and 4 … 20 mA
…GFxxx
frequency limit up to 30 kHz
…no declaration
standard version 1 kHz