9 | 17
5 Starting up
If an adjustment (A-K-1K / A-D-1K) has been ordered in combination with the amplifier (-s) and/or
transducers (-s) it may be necessary for a slight fine adjustment nonetheless. This is due to possible
various environmental influences as well as to mounting etc.
If any visual damage or malfunctions are noticed, the measuring system must be switched off and
marked appropriately.
■
Mounting transducer and amplifier
■
Connect transducer to the amplifier
■
Connect multimeter to analogue output
■
Connect supply voltage – consider pin assignment
■
Allow the system about 30 min. to warm up
■
Check function and calibration of the system
Notice:
The assignment of transducer / amplifier is to be complied with. After replacing a transducer
the calibration has to be checked.
It is to be noted that there is a slight dependence between zero point adjustment and amplification
(gain).