D000449-00 Rev. A
16
Starting and Completing a Workout Session
The procedures in this section describe:
•
Starting a workout session
•
Using the touchscreen features and controls
•
Adjusting Anti-Gravity Treadmill speed and direction
•
Adjusting pain levels
•
Displaying gait analytics
•
Starting live video monitoring
•
Using AlterG Assistant
Starting a Workout Session
Operator or User Action
Screen
1.
Tap START SESSION on the Welcome screen.
2.
Follow the instructions in the Before You Begin
screen, and then tap START CALIBRATION.
The calibration routine provides critical information
the Anti-Gravity Treadmill needs for accurate
control of body weight during exercise.
Note: For an accurate calibration, it is critical that
the user’s full body weight be applied to the belt
surface. Holding onto any other part of the
treadmill, for example, resting an arm on the top of
the enclosure while the treadmill calibrates, will
result in an erroneous calibration and inaccurate
body weight adjustment during exercise. It is
recommended that the user hold their arms out to
their sides or crossed on their chest, away from any
structures, during the calibration procedure.
When calibration is complete, the percentage of the
user’s body weight that can be unweighted is
displayed.
3.
Tap START SESSION. The workout session begins.