100-6262_R5 AP-4 User Guide Release Date: 4/28/20
8
5. If the Water or Concentrate show low (yellow bar) you will need to fill it prior to starting a cycle.
To add water, remove the
black cap labeled “WATER ONLY”. Fill with reverse osmosis (RO)
and/or deionized (DI) water until the display shows close to 100%.
Add 1 mL of Altacare
concentrate per gallon of water to assure freshness.
Make sure to add some water after
the concentrate is added to rinse it down the filler tube.
Do not
use tap water or filtered
drinking water.
Replace the cap.
6. To add Altacare Concentrate,
remove the black cap labeled “ALTACARE ONLY”.
IMPORTANT: Wear appropriate latex gloves and eye protection any time you are
handling the concentrated Altacare solution.
Carefully remove the cap from the Altacare bottle and pour into the fill spout. It is preferred to
wait until the Concentrate level is below 10% so that the entire contents of the bottle (1L) can
be added to avoid partial bottles needing to be stored. When the bottle is empty, rinse and
discard or recycle
7. If the Mixture shows EXPIRED, it will need to be drained before you can start a cycle. Press
the “Drain Mix Tank” button and allow it to drain into the drain bottle. Once it is fully drained, it
will reset the Mixture time to 72 hours.