Performing an Exam
Colorize
Colorize is the colorization of a conventional 2D mode image or Doppler Spectrum to enhance
the user's ability to discern 2D mode, M mode, and Doppler mode intensity valuations.
Colorize is NOT a Doppler mode. To adjust the colorize, rotate the
Colorize
soft key clockwise
or counter-clockwise.
Reject
Low echo information will not be displayed on the screen below the adjusted rejection level.
The rejection function determines the amplitude level below which echoes are suppressed
(rejected). Rejection set to high leads to bad tissue display.
To adjust the rejection values,
rotate the
Reject
soft key clockwise or counter-clockwise.
Power
Power controls the amount of acoustic power applied in all modes. To adjust the power, rotate
the
Power
soft key clockwise or counter-clockwise.
Edge Enhance
Edge Enhance demonstrates subtle tissue differences and boundaries by enhancing the gray
scale differences corresponding to the edges of structures. This function allows your system to
display the outline of tissues or an organ more clearly.
To adjust the edge enhance, rotate the
Edge Enhance
soft key clockwise or counter-clockwise.
Persist
Persistence provides a visible smoothing effect to the 2D-mode image by persisting lines
of image data for each frame of imaging. To adjust the persist, rotate the
Persist
soft key
clockwise or counter-clockwise.
Reverse
Reverse allows you to flip the image 180 degrees left/right. To flip the image horizontally,
press
the
key on the QWERTY keyboard.
Up/Down (If Preset)
Up/Down allows you to flip the image 180 degrees up/down. To flip the image vertically, press
the
U/D Flip
key on user-defined key.