![Alphasense OPC-N2 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/alphasense/opc-n2/opc-n2_user-manual_2906367034.webp)
072-0300
Alphasense User Manual
OPC-N2 Optical Particle Counter
Issue 5
Alphasense Ltd Page 34 of 34
December 2015
Sensor Technology House, 300 Avenue West, Skyline 120, Great Notley. Essex.CM77 7AA. UK
Tel: +44 (0) 1376 556700 - Fax: +44 (0) 1376 335899
Email: [email protected] - Web: www.alphasense.com
0x32
PM10 Byte2
0x32
PM10 Byte3
Save
Configuration
Variables in
non-volatile
memory
14/15
16/17
18
0x43
0x43
0xF3
5ms
Suggest that 10ms be
used as delay
between command
byte and following
byte.
0x3F
0x43
6us
Initial command byte
must be followed by
sequence of bytes
(shown in red).
0x3C
0x3F
“
0x3F
0x3C
“
0x3C
0x3F
“
0x43
0x3C
NA
Check Status 14/15
16/17
18
0xCF
0xCF
0xF3
NA
Enter
bootloader
mode
14/15
16/17
18
0x41
0x41
0xF3
NA
In general, suggest that the delay following first byte of any command sequence is 10ms
and the delay between subsequent byte transfers is 10us.
End of Manual