![Alphasense OPC-N2 Скачать руководство пользователя страница 33](http://html1.mh-extra.com/html/alphasense/opc-n2/opc-n2_user-manual_2906367033.webp)
072-0300
Alphasense User Manual
OPC-N2 Optical Particle Counter
Issue 5
Alphasense Ltd Page 33 of 34
December 2015
Sensor Technology House, 300 Avenue West, Skyline 120, Great Notley. Essex.CM77 7AA. UK
Tel: +44 (0) 1376 556700 - Fax: +44 (0) 1376 335899
Email: [email protected] - Web: www.alphasense.com
is unsigned 8bit
integer variable (1 or
0).
Write
Configuration
Variables 2
18
0x3B
0x3B
0xF3
Suggest that 10ms be
used as delay
between command
byte and following
byte.
AMSamplingInterv
alCount LSB
0x3B
AMSamplingIntervalC
ount is unsigned 16bit
integer variable.
AMSamplingInterv
alCount MSB
AMSamplingIntervalCount LSB
AMIdleIntervalCou
nt LSB
AMSamplingIntervalCount MSB
AMIdleIntervalCount
is unsigned 16bit
integer variable.
AMIdleIntervalCou
nt MSB
AMIdleIntervalCount LSB
AMFanOnInIdle
AMIdleIntervalCount MSB
AMFanOnInIdle is
unsigned 8bit integer
variable (1 or 0).
AMLaserOnInIdle AMFanOnInIdle
AMLaserOnInIdle is
unsigned 8bit integer
variable (1 or 0).
AMMaxDataArray
sInFile LSB
AMLaserOnInIdle
AMMaxDataArraysInF
ile is unsigned 16bit
integer variable.
AMMaxDataArray
sInFile MSB
AMMaxDataArraysInFile LSB
AMOnlySavePMD
ata
AMMaxDataArraysInFile MSB
AMOnlySavePMData
is unsigned 8bit
integer variable (1 or
0).
See note
0x30
0x30
0xF3
9ms
Output depends on
Firmware version
Read
histogram
data (and
reset
histogram)
0x30
Bin0 LSB
6us
See separate doc
Read PM data
(and reset
histogram)
18
0x32
0x32
0xF3
Suggest that 10ms be
used as delay
between command
byte and following
byte.
0x32
PM1 Byte0
PM1 is a float variable
occupying 4 bytes.
Units are ug/m
3
.
0x32
PM1 Byte1
0x32
PM1 Byte2
0x32
PM1 Byte3
0x32
PM2.5 Byte0
PM2.5 is a float
variable occupying 4
bytes. Units are
ug/m
3
.
0x32
PM2.5 Byte1
0x32
PM2.5 Byte2
0x32
PM2.5 Byte3
0x32
PM10 Byte0
PM10 is a float
variable occupying 4
bytes. Units are
ug/m
3
.
0x32
PM10 Byte1