Alpha MED-A64HE1 Скачать руководство пользователя страница 10

8

Stimulation Using Planar Microelectrodes

The MED64 Head Amplifier has a built-in 2 channel current-driven stimulator that enables stimulation via any of 
the 64 electrodes. The stimulus current amplitudes and channels are set in the [Acquire MED64R2 Data w/Stim] 
module in the MED64 Mobius software. Please refer to the Mobius Tutorial for setting up the stimulation 
parameters.

Fig.4 shows a schematic diagram of the extracellular electrical stimulation circuit for the MED64 system. The 
stimulus current flows from the selected electrode to reference electrodes. The applied stimulus current charges 
the “electrical double-layer capacitor” (Ce) at the interface of electrode, subsequently changing the Ce voltage 
(Vc). This current modulates the field potential, which results in hyperpolarization or depolarization of cellular 
membranes. The field potentials (Vf) change according to the output current (Is) as seen on the right in the 
Fig.4.  

Stimulus Artifacts and Biphasic Stimulation

During stimulation, signals (stimulus pulses) are recorded at recording channels (non-stimulated channels) due 
to crosstalk, causing "stimulus artifacts”. These stimulus artifacts can persist even after the stimulus pulse 
returns to zero and can interfere with the recording of real biological signals.

This is mainly due to following factors:

Fig.3.

  Mobius’ Control panel for setting stimulus parameter.

Select stimulator

Set

      Select stimulus electrode

s

stimulus waveform 

Ce:  Capacitance of electrical double-layer capacitor (electrode)
Re:  Resistance of electrode 
Rs:  Resistance of solution with slice
Is:   Stimulator output current 
Vs:  Stimulator output voltage 
Vc:  Voltage of Ce 
Vf:   Voltage of Rs (Voltage change in the field)

Stimulator output

current (Is)

Stimulator output

voltage (Vs)

Voltage of Ce (Vc)

Voltage of Rs (Vf)
(Changes in the 

 field potentials)

Fig.4

. Stimulation using microelectrodes at the MED64 system.

Содержание MED-A64HE1

Страница 1: ...Product Manual MED64 Head Amplifier 64 CHANNEL HEAD AMPLIFIER P N MED A64HE1 A low noise multi electrode array system for in vitro electrophysiology...

Страница 2: ...mages resulting from the use of information contained in this document or from the use of programs and source code that may accompany it In no event shall the publisher and the author be liable for an...

Страница 3: ...e MED64 Main Amplifier to other Components 6 Signal Acquisition with the MED64 system 7 Amplifier Input Impedance Requirements for avoiding Signal Attenuation 7 Stimulation Using Planar Microelectrode...

Страница 4: ...ower cords with cuts scratches or poor connectors as this may result in fire or electric shock Excessive bending pulling or slicing of the cord should also be avoided Do not pull on the cord when you...

Страница 5: ...symbols used on equipment and documentation Environmental Conditions Indoor use Altitude up to 2000 m Temperature 5 40 C Maximum relative humidity 80 for temperatures up to 31 C decreasing linearly to...

Страница 6: ...livered via any of the 64 electrodes using the built in stimulator in the MED64 Head Amplifier Components and their functions Front panel 1 POWER Turns amplifier ON and OFF 2 STIMULUS CURRENT Selectio...

Страница 7: ...2 STIMULUS OUTPUT on the 64 CHANNEL MAIN AMPLIFIER MED A64MD1 5 CONTROL INPUT Input for control of stimulus channels Connects to the CONTROL OUTPUT on the 64 CHANNEL MAIN AMPLIFIER MED A64MD1 6 SIGNAL...

Страница 8: ...idified incubator It is connected with a cable as long as 2m without noise or signal attenuation The MED64 system does NOT usually require a Faraday cage or vibration isolation table necessary for con...

Страница 9: ...have very low impedances at 1kHz 10 k for 50 m electrodes 40 k for 20 m electrodes The impedances are small even at low frequencies as seen in Fig 2 Thus the input impedance of 100 M for the MED64 Hea...

Страница 10: ...field potentials Vf change according to the output current Is as seen on the right in the Fig 4 Stimulus Artifacts and Biphasic Stimulation During stimulation signals stimulus pulses are recorded at...

Страница 11: ...ositive components as seen in Fig 5 When biphasic stimulation is applied the electrode is charged and discharged quickly However when mono phasic stimulation is applied the electrodes need longer time...

Страница 12: ...s 50 000 pF for 50 x 50 m electrodes 22 000 pF for 20 x 20 m electrodes allowing stimulation with current amplitudes up to 200 A for 0 2 msec duration for each phase is 0 1 msec DO NOT use stimulus cu...

Страница 13: ...n maladjustment of user controls improper maintenance modifications or service by anyone other than AMS or damage attributable to acts of God Specifications Amplifier Stimulator General Power supply u...

Страница 14: ...ntific Inc Saito Bio Incubator 209 7 7 15 Saito asagi Ibaraki Osaka 567 0085 Japan Phone 81 72 648 7973 FAX 81 72 648 7974 http www med64 com support med64 com Manufactured by Alpha MED Scientific Inc...

Отзывы: