LED Indicators 

 

• AC Power with Battery Backup - Solid Green 
• AC Power with Low Battery or No Battery – Solid Red 
• Battery Only - Blinks Green 
• Battery Only (Low Battery) – Blinks Red 
 

Note:

 KT8P0 operates on battery only and will blink on time every minute to conserve battery life. 

 
 

User Calibration 

 

Your instrument was carefully tested and calibrated before being shipped from the factory. Only the Replaceable Sensor is 
calibrated.  The sensor, and all calibration defaults and adjustments are stored on the Replaceable Sensor.  Accuracy is strictly 
controlled by the sensor 

– no adjustments are made by the Instrument.  Readings are sent from the Replaceable Sensor to the 

Instrument for storage purposes only.  Certificates of Calibration / NISTs are only supplied for the Replaceable Sensor for this 
reason.   When a Replaceable Sensor is due for recalibration the existing Replaceable Sensor can be replaced with a new 
Replaceable Sensor thus eliminating the need to return the entire instrument for recalibration.  Replaceable Sensors can be 
returned for recalibration as well.  For greatest accuracy, we recommend replacing the sensor every 6-12 months.  For re-
calibration, call customer service at (630) 543-3747. If you wish to do calibration yourself, follow these procedures: 
 

1.

 

To activate calibration mode, turn the unit on and press and hold both the 

On/Off 

button and the 

Adjust-Down

 button 

(making sure to press the 

Adjust-Down

 button first so the unit will not turn off). The led will blink in amber rapidly for about 

five seconds, and turn solid green. Release the 

On/Off

 button and the 

Adjust-Down

 button, the led will then blink amber 

at which point only the pen being adjusted will show on the display. 
 

2.

 

To raise the unit of measurement, press the 

Adjust-Up

 button. To lower the unit of measurement, press the 

Adjust-

Down

 button. 

Pen home

 switches between the red and blue pens on a two pen unit and stores the current adjustment 

value.  
 

3.

 

When calibration is complete, press the 

On/Off

 button. Calibration is stored in memory even after you turn the unit off. 

User calibration information will not be lost if AC power fails. 
 

Note: 

After two hours, if no buttons are pressed, the unit will time out of user calibration and resume normal operation. The 

user calibration is stored separately from the factory calibration. If you wish to cancel out your user calibration, simply enter 
calibration mode and toggle through the steps without adjusting displayed readings. Exit by pressing the On/Off button. 
You have now restored factory calibration. 
 

It is recommended that you use a controlled chamber when calibrating temperature and humidity. Calibrating in 
an open room is not recommended as humidity can vary greatly within a very small area.

 

 
 

Need to Know 

 
Pen adjust may be needed if the pen(s) and display do not match.  While the unit is on, press and hold both the 

Pen Home

  

button and the 

On/Off 

button (making sure to press the 

Pen Home

 button first so the unit will not turn off). The led will blink 

rapidly between amber and green for about five seconds. The LED will then turn green. Release the 

Pen Home 

and 

On/Off  

buttons. The LED will flash amber and green for one second and the LED will turn off. 
 

One-Pen Models: 

 

The red pen will move to the outer edge of the chart and is ready to be adjusted. Use the 

Adjust-Up

 and 

Adjust-Down 

buttons to move the pen to the maximum temperature line on the chart. Once the pen is set, the Pen Home button will move 
the pen to reaffirm the new pen position. Press the 

On/Off 

button to exit Pen Home Adjust. 

 

Two-Pen Models:  

Both pens will move to the outer edge of the chart, and then the shorter red pen will move to the current position. Use the  

Adjust-Up 

and 

Adjust-Down

 buttons to move the pen to the maximum temperature line on the chart. Once the shorter red 

pen is set, the Pen Home button will move the shorter red pen to the outer edge of the chart and the longer blue pen will move  
 
 

 

 

 