Alpha Omega Instruments OXY-SEN Скачать руководство пользователя страница 6

ALPHA OMEGA INSTRUMENTS CORP.
OXY-SEN OXYGEN MONITOR

         Instruction Manual

SECTION 3.0

INSTALLATION PROCEDURES

Unpacking the Instrument

Upon opening the shipping container, carefully unpack the monitor to check if the outer surfaces
have been damaged. If so, report the findings immediately to Alpha Omega Instruments who will
provide further instructions. If there is no apparent damage, check the contents 
to ensure all items were shipped.  In some cases, items may be back-ordered.  

All damage and shortage claims must be made 
known to Alpha Omega Instruments within 10 
days after receipt of shipment.

Mounting the 

OXY-SEN Oxygen Monitor

The mounting slots for the monitor electronics are 2-5/8" apart vertically, and  5-3/8 " apart 
horizontally.   The slots can accommodate # 6 through #8 screws.

Powering the OXY-SEN Oxygen Monitor A/C Power Source Version 

The 

OXY-SEN Oxygen Monitor

 is powered from 115 VAC/60 HZ. or 208/240 VAC, Hz., at 10 

ma. MAX.   The Oxy-Sen Oxygen  Monitor  senses the input voltage and adjusts automatically 
for proper operation.

Powering the OXY-SEN Oxygen Monitor 24 VDC Power Source Version

The OXY-SEN Oxygen Monitor is powered from a 24 volt DC power source. The unit is 
internally fused for 500 mA. All manual references to AC power input do not apply. 

DO NOT 

CONNECT THE 24VDC POWER SOURCE VERSION TO ANY AC (115/220 VAC) POWER 
SOURCE!

Mounting the 

Oxy-Sen Oxygen Sensor

The pre-cabled sensor mounts into a tapped hole with a M16X1 thread.  Insure that only
finger tight pressure is applied to the sensor when mounting, as the o-ring supplied with 
the sensor will supply a good seal with finger tight torque to the sensor.

The sensor connects to the monitor's electronics with a pre-wired cable connector, 
installed into the connector labeled "Sensor" (use the dimple on the connector for 
alignment.

Connecting the Optional Analog Output

 

Cable

When ordered with the optional analog output, the monitor is equipped with a 10 foot (3 
meter) cable. This cable can be used to access either the 0-25% or 0-100% oxygen 
output range. The location of these connections is shown in Figure 1.0 on page 7.

3

Содержание OXY-SEN

Страница 1: ...e 100 Lincoln RI 02865 Phone 401 333 8580 Fax 401 333 5550 Website www aoi corp com Email salescontact aoi corp com VERSION 2 2 COPYRIGHT 06 18 ALPHA OMEGA INSTRUMENTS CORP ALL RIGHTS RESERVED INCLUDING THE RIGHT TO REPRODUCE THIS MANUAL OR ANY PORTION THEREOF IN ANY FORM ...

Страница 2: ...RRANTED AS BEING FIT FOR A PARTICULAR PURPOSE AND THERE IS NO WARRANTY OF MERCHANTABILITY This warranty applies only if i the items are used solely under the operating conditions and manner recommended in this manual specifications or other literature ii the items have not been misused or abused in any manner or repairs attempted thereon iii written notice of the failure within the warranty period...

Страница 3: ...N 2 0 SYSTEM DESCRIPTION 2 SECTION 3 0 INSTALLATION PROCEDURES 3 SECTION 4 0 OPERATING PROCEDURES 4 SECTION 5 0 CALIBRATION PROCEDURES 5 SECTION 6 0 REPLACEMENT OF OXYGEN SENSOR 6 FIGURE 1 0 OXY SEN OXYGEN MONITOR DIMENSIONAL DRAWING 7 FIGURE 2 0 OXY SEN OXYGEN MONITOR ANALOG INTERFACE CABLE 8 APPENDIX A MSDS FOR THE 2SEN OXYGEN SENSOR 9 ii ...

Страница 4: ...ics ELECTRICAL Display 3 1 2 digit liquid crystal Input Power 115 230 VAC 50 60Hz or 24 VDC Analog Output Optional 4 20 mADC scaled over 0 100 O2 and 0 25 O2 selectable by connector position SAMPLE GAS CHARACTERISTICS Storage Temperature Range 40 to 104 F 5 to 40 C Sample Gas Pressure Limits 2 psig 0 14 kg cm2 CONSTRUCTION Dimensions of Electronics Enclosure 5 3 4 in 146 05 mm Width 5 1 2 in 139 7...

Страница 5: ...brane is reduced electrochemically at the gold electrode A resistor and a thermistor for temperature compensation are connected between the cathode and anode As a result the lead oxygen battery is always discharged Current flowing through the resistor and thermistor is proportional to the oxygen concentration of the gas in contact with the Teflon membrane By measuring the voltage between the resis...

Страница 6: ...Oxygen Monitor senses the input voltage and adjusts automatically for proper operation Powering the OXY SEN Oxygen Monitor 24 VDC Power Source Version The OXY SEN Oxygen Monitor is powered from a 24 volt DC power source The unit is internally fused for 500 mA All manual references to AC power input do not apply DO NOT CONNECT THE 24VDC POWER SOURCE VERSION TO ANY AC 115 220 VAC POWER SOURCE Mounti...

Страница 7: ...mbers etc where such threaded holes can be made quite easily Once the sensor has been properly installed and the O ring properly seated then the other end of the sensor cable may be connected to the bottom of the electronics labeled Sensor Alpha Omega Instruments does offer an optional T like flow through gas housing for the OXY SEN If this option has been ordered the housing will be equipped with...

Страница 8: ...libration of the OXY SEN every 2 3 months Calibration Gas The oxygen sensor used in the OXY SEN Oxygen Monitor has a linear output As a result it can be calibrated using a source of fresh air which contains 20 9 O2 For gas measurement expected to be over 20 9 a calibration gas at or above the anticipated measurements may improve calibration accuracy An optional calibration fixture P N OS 1CAL is a...

Страница 9: ...the instrument 2 0 Locate the sensor cable located on the bottom of the enclosure and unplug the sensor 3 0 Locate the oxygen sensor where it is mounted and remove the sensor by unscrewing the sensor by rotating in a counterclockwise motion Discard the old sensor see precautionary note below and install the replacement sensor being careful not to over tighten the sensor The sensor has an O ring se...

Страница 10: ...ALPHA OMEGA INSTRUMENTS CORP OXY SEN OXYGEN MONITOR Instruction Manual Figure 1 0 Dimensional Drawing 7 ...

Страница 11: ...ALPHA OMEGA INSTRUMENTS CORP OXY SEN OXYGEN MONITOR Instruction Manual Figure 2 0 Analogue Output Cable The cable connector is a 4 Pin mini DIN Connector AMP P N 171 2604 8 ...

Страница 12: ...LV 10 PPM NOTE TLV and PEL are for concentrated 90 100 Acetic Acid actual solution is less than 50 Lead CAS 7439 92 1 OSHA PEL 0 03mg m3 SECTION III PHYSICAL CHEMICAL CHARACTERISTICS Boiling Point Not Available Specific Gravity Not Available Vapor Pressure Not Available Vapor Density Not Available Evaporation Rate Not Available Physical State Liquid Melting Point Not Available pH 3 5 7 0 Flash Poi...

Страница 13: ...Reproductive Effects None identified Effects of Overexposure INHALATION Tightness and pain in chest coughing difficult breathing SKIN CONTACT Irritation EYE CONTACT Irritation SKIN ABSORPTION May be harmful INGESTION Is harmful and may be fatal headache nausea vomiting dizziness gastrointestinal irritation CHRONIC EFFECTS Anemia kidney damage blurred vision lead build up in the central nervous sys...

Страница 14: ...Severe irritation or burns of respiratory system SKIN CONTACT Severe burns may cause dermatitis EYE CONTACT Severe burns permanent eye damage SKIN ABSORPTION None identified INGESTION Burns to mouth and throat nausea vomiting gastrointestinal irritation diarrhea shock may be fatal CHRONIC EFFECTS Lung damage teeth damage Target Organs Respiratory system eyes skin teeth lungs Medical Conditions Gen...

Страница 15: ... gloves are recommended Protective measures during cell replacement Before opening the packaging containing the sensor cell check the sensor cell for leakage If the sensor cell leaks do not open the container If there is liquid around the cell while in the instrument use the protection listed above in this section SECTION VIII STORAGE AND HANDLING PRECAUTIONS Storage Requirements Store in a cool w...

Отзывы: