Alpha Omega Instruments OXY-SEN Скачать руководство пользователя страница 5

ALPHA OMEGA INSTRUMENTS CORP.
OXY-SEN OXYGEN MONITOR

         Instruction Manual

SECTION 2.0

SYSTEM

 

DESCRIPTION

General Description

The OXY-SEN Oxygen Monitor is a percent oxygen monitor designed to provide accurate and 
dependable percent oxygen measurements in a variety of gases.  The monitor enclosure is 
made from aluminum, and is rated for general purpose service.  The monitor is powered from 
115 VAC, 208/240 VAC  (auto-adjusts to input voltage)  or 24 VDC (must be selected at time of 
order).  

Ambient Temperature Electrochemical Sensor with EES

The OXY-SEN Oxygen Monitor features an extended life oxygen sensor with EES (enhanced 
electrolyte system) a weak acid electrolyte system.  The weak acid electrolyte system retards 
passivation of the sensor anode by allowing the products of oxidation to dissolve in the acid 
electrolyte.  In effect, the sensor is renewed continuously as the weak acid electrolyte tolerates 
over 20 times the lead oxide (PbO) than potassium hydroxide (KOH) based sensors. The result 
is a sensor with a greatly extended useful life.

The extended life sensor is a lead-oxygen battery comprised of a lead anode, a gold cathode, 
and a weak acid electrolyte.  A gold electrode is bonded onto a non-porous Teflon (FEP) 
membrane.  A small amount of oxygen permeating through the membrane is reduced 
electrochemically at the gold electrode. A resistor and a thermistor (for temperature 
compensation) are connected between the cathode and anode.  As a result, the lead-oxygen 
battery is always discharged.

Current flowing through the resistor and thermistor is proportional to the oxygen concentration of
the gas in contact with the Teflon membrane.  By measuring the voltage between the resistor 
and the thermistor, oxygen concentration can be accurately determined. The two electrode 
reactions are depicted below:

Cathode: O

2

 + 4H

+

 + 4e                

2H

2

O

Anode:   2Pb + 2H

2

O                      

 2PbO + 4H

+

 + 4e

Overall:  O

+ 2Pb

2PbO

If PbO generated at the anode remains on the anode, the anode's potential will change and the 
output voltage of the sensor would become extremely low. However, in actuality the surface of 
the lead anode is continuously renewed because PbO is dissolved back into the electrolyte.
There is a maximum level to the solubility of PbO in the weak acid electrolyte.  As a result, when
the electrolyte becomes saturated with PbO, the sensitivity of the sensor will begin to drop 
signaling time for replacement of the sensor. When the extended life oxygen sensor is used in 
ambient air monitoring applications, carbon dioxide (CO

2

) that is present will not adversely affect

the sensor's performance.  Due to the weak acid electrolyte, CO

2

 will not react with the 

electrolyte to form potassium carbonate as it does with KOH based sensors.  As a result, there 
is no loss in output due to the presence of CO

2

.

2

Содержание OXY-SEN

Страница 1: ...e 100 Lincoln RI 02865 Phone 401 333 8580 Fax 401 333 5550 Website www aoi corp com Email salescontact aoi corp com VERSION 2 2 COPYRIGHT 06 18 ALPHA OMEGA INSTRUMENTS CORP ALL RIGHTS RESERVED INCLUDING THE RIGHT TO REPRODUCE THIS MANUAL OR ANY PORTION THEREOF IN ANY FORM ...

Страница 2: ...RRANTED AS BEING FIT FOR A PARTICULAR PURPOSE AND THERE IS NO WARRANTY OF MERCHANTABILITY This warranty applies only if i the items are used solely under the operating conditions and manner recommended in this manual specifications or other literature ii the items have not been misused or abused in any manner or repairs attempted thereon iii written notice of the failure within the warranty period...

Страница 3: ...N 2 0 SYSTEM DESCRIPTION 2 SECTION 3 0 INSTALLATION PROCEDURES 3 SECTION 4 0 OPERATING PROCEDURES 4 SECTION 5 0 CALIBRATION PROCEDURES 5 SECTION 6 0 REPLACEMENT OF OXYGEN SENSOR 6 FIGURE 1 0 OXY SEN OXYGEN MONITOR DIMENSIONAL DRAWING 7 FIGURE 2 0 OXY SEN OXYGEN MONITOR ANALOG INTERFACE CABLE 8 APPENDIX A MSDS FOR THE 2SEN OXYGEN SENSOR 9 ii ...

Страница 4: ...ics ELECTRICAL Display 3 1 2 digit liquid crystal Input Power 115 230 VAC 50 60Hz or 24 VDC Analog Output Optional 4 20 mADC scaled over 0 100 O2 and 0 25 O2 selectable by connector position SAMPLE GAS CHARACTERISTICS Storage Temperature Range 40 to 104 F 5 to 40 C Sample Gas Pressure Limits 2 psig 0 14 kg cm2 CONSTRUCTION Dimensions of Electronics Enclosure 5 3 4 in 146 05 mm Width 5 1 2 in 139 7...

Страница 5: ...brane is reduced electrochemically at the gold electrode A resistor and a thermistor for temperature compensation are connected between the cathode and anode As a result the lead oxygen battery is always discharged Current flowing through the resistor and thermistor is proportional to the oxygen concentration of the gas in contact with the Teflon membrane By measuring the voltage between the resis...

Страница 6: ...Oxygen Monitor senses the input voltage and adjusts automatically for proper operation Powering the OXY SEN Oxygen Monitor 24 VDC Power Source Version The OXY SEN Oxygen Monitor is powered from a 24 volt DC power source The unit is internally fused for 500 mA All manual references to AC power input do not apply DO NOT CONNECT THE 24VDC POWER SOURCE VERSION TO ANY AC 115 220 VAC POWER SOURCE Mounti...

Страница 7: ...mbers etc where such threaded holes can be made quite easily Once the sensor has been properly installed and the O ring properly seated then the other end of the sensor cable may be connected to the bottom of the electronics labeled Sensor Alpha Omega Instruments does offer an optional T like flow through gas housing for the OXY SEN If this option has been ordered the housing will be equipped with...

Страница 8: ...libration of the OXY SEN every 2 3 months Calibration Gas The oxygen sensor used in the OXY SEN Oxygen Monitor has a linear output As a result it can be calibrated using a source of fresh air which contains 20 9 O2 For gas measurement expected to be over 20 9 a calibration gas at or above the anticipated measurements may improve calibration accuracy An optional calibration fixture P N OS 1CAL is a...

Страница 9: ...the instrument 2 0 Locate the sensor cable located on the bottom of the enclosure and unplug the sensor 3 0 Locate the oxygen sensor where it is mounted and remove the sensor by unscrewing the sensor by rotating in a counterclockwise motion Discard the old sensor see precautionary note below and install the replacement sensor being careful not to over tighten the sensor The sensor has an O ring se...

Страница 10: ...ALPHA OMEGA INSTRUMENTS CORP OXY SEN OXYGEN MONITOR Instruction Manual Figure 1 0 Dimensional Drawing 7 ...

Страница 11: ...ALPHA OMEGA INSTRUMENTS CORP OXY SEN OXYGEN MONITOR Instruction Manual Figure 2 0 Analogue Output Cable The cable connector is a 4 Pin mini DIN Connector AMP P N 171 2604 8 ...

Страница 12: ...LV 10 PPM NOTE TLV and PEL are for concentrated 90 100 Acetic Acid actual solution is less than 50 Lead CAS 7439 92 1 OSHA PEL 0 03mg m3 SECTION III PHYSICAL CHEMICAL CHARACTERISTICS Boiling Point Not Available Specific Gravity Not Available Vapor Pressure Not Available Vapor Density Not Available Evaporation Rate Not Available Physical State Liquid Melting Point Not Available pH 3 5 7 0 Flash Poi...

Страница 13: ...Reproductive Effects None identified Effects of Overexposure INHALATION Tightness and pain in chest coughing difficult breathing SKIN CONTACT Irritation EYE CONTACT Irritation SKIN ABSORPTION May be harmful INGESTION Is harmful and may be fatal headache nausea vomiting dizziness gastrointestinal irritation CHRONIC EFFECTS Anemia kidney damage blurred vision lead build up in the central nervous sys...

Страница 14: ...Severe irritation or burns of respiratory system SKIN CONTACT Severe burns may cause dermatitis EYE CONTACT Severe burns permanent eye damage SKIN ABSORPTION None identified INGESTION Burns to mouth and throat nausea vomiting gastrointestinal irritation diarrhea shock may be fatal CHRONIC EFFECTS Lung damage teeth damage Target Organs Respiratory system eyes skin teeth lungs Medical Conditions Gen...

Страница 15: ... gloves are recommended Protective measures during cell replacement Before opening the packaging containing the sensor cell check the sensor cell for leakage If the sensor cell leaks do not open the container If there is liquid around the cell while in the instrument use the protection listed above in this section SECTION VIII STORAGE AND HANDLING PRECAUTIONS Storage Requirements Store in a cool w...

Отзывы: