Alpha Omega Instruments 9510 Series Скачать руководство пользователя страница 22

3.3.3

Fail Safe

Switches #1 through #4 control the fail-safe operation for each of the four alarm relays. Each switch must
be turned "ON" to be fail-safe (refer to FIGURE 3.0).  In the fail-safe mode,  the normal or non-alarm state
of each of the four (4) alarm relays is energized.  When an alarm signal is generated, the respective relay
for the alarm is de-energized.  This would be the same relay state (de-energized) that would be achieved
if the power source (AC, DC, or Battery) were interrupted.  The factory default setting for switches #1
through #4 is "OFF" (normal operation or non fail-safe). 

 Please remember that when Fail Safe

Behavior is selected for the relays, the drain from the optional  NiMH battery back-up is
approximately doubled, and the analyzer operating time is approximately halved.

3.3.4

Manual Clear / Auto Clear

There are two possible conditions that effect the way the instrument will respond to alarm cancellation.
One is Auto-Clear operation and the other is Manual Clear. The mode of operation is determined by the
setting switch #6 of SW6.

When switch 6 of DIP Switch 6 is placed in the “Off”  or ‘”Manually Clear Al arms” position, the user is
required to manually clear the alarm indication from the analyzer even if the original alarm condition has
returned to a non-alarm level.  When the switch is in the “On” or “Auto-Clear Alarms” position, the analyzer
will automatically clear the three (3) alarm indications (LED, Audible Alarm, and Relay) when the original
alarm condition returns to a non-alarm level.

3.3.4.1 Manual Clear Operation

In the Manual Clear mode (SW6, switch #6 is  "OFF), the analyzer will not automatically clear the three (3)
alarm indicators.  The user must first cancel the audible alarm by pushing the appropriate alarm button.
Only after the audible alarm has been canceled can the user clear the alarm condition by again pushing
the appropriate alarm button.  If the set-point is to remain the same, simply push the appropriate alarm
button a third time. 

Note:

  If the audible alarm is activated, pressing any alarm button that is not associated with an alarm

condition will temporarily quiet the audible alarm.  Upon exiting the alarm set condition or if the monitor
times out because of no user input, the audible alarm will resume awaiting a manual clear operation as
described above

3.3.4.2 Auto Clear Operation

In the Auto Clear mode (SW6, switch #6 is ON), the Series 9510 Carbon Dioxide Analyzer will
automatically reset the 3 alarm indications mentioned above when the alarm condition clears.   Under
Auto-Clear operation, the silenced audible alarm may automatically come back on if the  CO

2

 reading

should go out of alarm range and then back into alarm condition. Also, after silencing the audible alarm, all
alarms must be cleared to re-enable the audible alarm.

3.3.5

Setting the Alarm Levels

Assuming that no alarms are currently activated (no LEDs are lit), press the desired alarm switch on the
front panel, "Alarm 1", "Alarm 2", or "Alarm 3" . The numerical value in the LCD is the existing alarm value
associated with that alarm channel.  When the alarm switch is pressed, the LED directly above the switch
will light indicating that channel is in the alarm set mode.   Any of the three CO

2

 Alarms can be changed

from low to high level or vice versa by pressing the UP and DOWN arrows simultaneously.    To set the
CO

2

 alarm values, press the DOWN arrow to lower the CO

2

 alarm value or the UP arrow to increase the

value. The longer either arrow is held down, the more rapidly the alarm values will scroll in the display.
When the value in the display is close to the desired  CO

2

 set point value, it is recommended that pressure

be released from the switch. To obtain the final value, apply momentary pressure to the switch to change
values in small increments.  When finished setting the alarm, press the associated alarm switch.  The LED

ALPHA OMEGA INSTRUMENTS CORP.
SERIES 9510 CARBON DIOXIDE ANALYZER                                                   INSTRUCTION MANUAL

16

Содержание 9510 Series

Страница 1: ...Rd Suite 100 Lincoln RI 02865 Tel 401 333 8580 Fax 401 333 5550 Email contact aoi corp com VER 1 3 11 06 SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL ALPHA OMEGA INSTRUMENTS salescontact aoi...

Страница 2: ...ommended in this manual specifications or other literature ii the items have not been misused or abused in any manner or repairs attempted thereon iii written notice of the failure within the warranty...

Страница 3: ...ALLING OR REMOVING AC POWER OR DATA SIGNAL CONNECTIONS AND WHEN PERFORMING ANY WORK INSIDE THE INSTRUMENT ENCLOSURE RF Disclaimer This instrument generates and uses small amounts of radio frequency en...

Страница 4: ...mple Pump 2 1 2 8 Coalescing Filter 2 1 2 9 Horn Strobe 2 1 3 Non Dispersive Infrared Sensor for Carbon Dioxide 3 1 4 General Specifications 5 SECTION 2 0 INSTALLATION PROCEDURES 6 2 1 Unpacking the I...

Страница 5: ...rm Processing 15 3 3 2 User Settings and Alarm Defaults 15 3 3 3 Fail Safe 16 3 3 4 Manual Clear Auto Clear 16 3 3 4 1 Manual Clear Operation 16 3 3 4 2 Auto Clear Operation 16 3 3 5 Setting the Alarm...

Страница 6: ...OL COMMANDS WITH RS 485 28 7 1 General 28 7 2 Special Commands for RS 485 29 7 3 Added RS 485 Commands 30 7 4 RS 485 Enhanced Commands 32 APPENDICES Appendix A Bench Top Portable Configuration 33 Part...

Страница 7: ...h two standard analog outputs 4 20 mADC and 0 2 VDC which are scalable by the user For communications capability the Series 9510 Carbon Dioxide Analyzer can be equipped with optional RS 232C serial co...

Страница 8: ...r several analyzers and a host system over the same communications channel The RS 485 format allows both sending and receiving of signals over greater distances than RS 232C making it ideal for instal...

Страница 9: ...tween the IR radiation that reaches the detector with and without the target gas in the gas cell The absorption is calculated for the infrared signals measured under zero gas gas that does not have in...

Страница 10: ...factor in traditional IR sensors is the temperature dependence of the detector filter pair The use of a single detector and single optical filter in calculating the absorption ratio in Alpha Omega Ins...

Страница 11: ...CO2 Alarm Relays Three 3 SPDT Form C contacts rated 10 A 30V DC 115 230 VAC Alarms may be cleared manually or automatically by user selection Instrument Status Alarm Relay 1 SPDT Form C contact rated...

Страница 12: ...TRICAL INSTALLATION SHOULD BE PERFORMED BY A LICENSED ELECTRICIAN AND SHOULD COMPLY WITH APPLICABLE FEDERAL STATE OR LOCAL ELECTRICAL SAFETY CODES NOTE For Explosion Proof Systems please refer to Appe...

Страница 13: ...MAIN PRINTED CIRCUIT BOARD FIGURE 1 0 ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 7 SW 6 Fuse...

Страница 14: ...larms defaulting to factory setting of low CO2 alarms To configure any of the three CO2 alarm relays to act as low alarms please refer to Section 3 3 Alarm 4 is the instrument status alarm which is no...

Страница 15: ...COM 19 NC to 17 COM Low Battery or Instrument Status Relay 4 21 NO to 20 COM 22 NC to 20 COM High or Low CO2 Alarm3 Relay 3 24 NO to 23 COM 25 NC to 23 COM High or Low CO2 Alarm2 Relay 2 27 NO to 26...

Страница 16: ...rminals 12 and 13 it must be removed prior to using the 4 20 mADC output NOTE IF THE 4 20MA DC OUTPUT IS NOT GOING TO BE USED IT SHOULD BE SHORTED TO AVOID AN OPEN LOOP WARNING VIA TERMINAL 10 B LO 2...

Страница 17: ...n WK4041BK Changing the AC input voltage can be done as follows 1 Turn Power off and disconnect the Series 9510 Carbon Dioxide Analyzer from all AC power 2 Swing out the front panel to gain access to...

Страница 18: ...operation on a fully charged set of batteries without pump With a pump one may expect to obtain up to 3 4 hours of continuous operation before charging of the batteries will be necessary The batterie...

Страница 19: ...d Indicators The front panel of the Series 9510 Carbon Dioxide Analyzer contains a 4 1 2 digit liquid crystal display LCD three 3 alarm set membrane switches three 3 alarm LEDs and up and down membran...

Страница 20: ...w keys until the high value desired is displayed There are a few shortcuts for entering new values Pressing the A1 key sets the value to 0 Pressing the A2 key sets the value to mid scale for the range...

Страница 21: ...WITCH SETTINGS FIGURE 3 0 3 3 2 Default User Settings With switch 8 of SW6 in the OFF position the Series 9510 Carbon Dioxide Analyzer uses the settings of switches 1 7 for instrument alarm configurat...

Страница 22: ...larm button If the set point is to remain the same simply push the appropriate alarm button a third time Note If the audible alarm is activated pressing any alarm button that is not associated with an...

Страница 23: ...odel Number 9PRR or 9LPR Flow rates through the analyzer should be between 100 to 300 milliliters per minute ml min for precent systems and approximately 150 millimeters a minute for trace systems If...

Страница 24: ...elivering a gas sample at a pressure at under 1 0 PSIG Connect the calibration gas to the fitting on the rear of the monitor labeled INPUT 3 Power up the Series 9510 Carbon Dioxide Analyzer as describ...

Страница 25: ...uick connect gas fitting suitable for rigid plastic tubing This fitting is located either on the rear or bottom of the instrument and is labeled Calibration Gas Exercising the same precautionary measu...

Страница 26: ...tandoffs which hold the CO2 sensor in place and lift up the sensor so that the battery housing cover can be removed 4 Remove the four 4 screws that secure the battery cover in place and remove the cov...

Страница 27: ...wn within brackets Command Description Aa bb b L H Enter Alarm set Bccccc Enter Baud rate select C bb b Enter Calibrate user to enter cal value D string Enter Disable Security E string Enter Enable Se...

Страница 28: ...garbled data output Now you must change your terminal s baud rate and reestablish communication by pressing Enter C Command Calibrate to known calibration standard i e 40 ppm Using a calibration gas...

Страница 29: ...nd 3 H Command Help Screen Displays a help screen H Enter M Command Manually clear all alarms toggle This command toggles between M anual and Automatic clearing of alarms The clearing of an alarm is s...

Страница 30: ...ere Alarm 2 and Alarm 3 will automatically shut off This could tell the pump to shut off and allow for the window to be opened by Alarm 1 Eventually due to some chemical reaction in the room the CO2 l...

Страница 31: ...Relay 3 Energized Conditions AC inp ok 4 20mA ok Open Collector output off Batt ok 22 Aux Relay De energized Alarms to be cleared MANUALLY Signal Mode Commands that affect the output of the V command...

Страница 32: ...Connect the Series 9510 Carbon Dioxide Analyzer to a terminal or computer as follows Series 9510 Computer GNDRXDTXD R1 14 15 16 RXD TXD GND ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANAL...

Страница 33: ...CO2 for alarm setting L H Optionally set to L ow or H igh alarm ccccc Baud rate number from 150 to 9600 d Number designating Relay 1 to 4 ON OFF 1 0 ON is the same as 1 etc name String for accessing m...

Страница 34: ...with the B command for setting the baud rate This is a handy little time saver For instance to set all units to 9600 bps just type GB9600 Or in other words Global Baud 9600 Make sure you change your t...

Страница 35: ...alarm signal No other commands are available for G lobal setting 7 3 Added RS485 Commands L Command Local name assignment This command allows the user to name each Series 9510 that might be on the RS...

Страница 36: ...up on the display This means you have successfully performed a Factory boot The instrument will then proceed to a normal warm boot to read the switch settings etc At this time you can name the second...

Страница 37: ...y Q uiet any unit by following the command with a valid unit name S Command Signal Mode select Same enhancement as above V Command View settings This has the same enhancement as the M Q and S commands...

Страница 38: ...APPENDIX A BENCH TOP PORTABLE CONFIGURATION PART NUMBER BTP ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 32...

Страница 39: ...APPENDIX B PANEL MOUNTING CONFIGURATION PART NUMBERS PNL PNR PTX FRONT VIEW ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 33...

Страница 40: ...APPENDIX B PANEL MOUNTING CONFIGURATION PART NUMBERS PNL PNR PTX SIDE VIEW ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 34...

Страница 41: ...APPENDIX B PANEL MOUNTING CONFIGURATION PART NUMBERS PNL PNR PTX REAR VIEW ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 35...

Страница 42: ...APPENDIX B PANEL MOUNTING CONFIGURATION PART NUMBERS PNL PNR PTX PANEL CUT OUT DETAILS ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 36...

Страница 43: ...X FRONT VIEW Note The item shown above is an explosion proof oxygen measuring system The Series 9510 Carbon Dioxide Analyzer is identical to that shown above in terms of mechanical layout ALPHA OMEGA...

Страница 44: ...ystem The Series 9510 Carbon Dioxide Analyzer is identical to that shown above in terms of mechanical layout Due to options ordered with the Series 9510 Carbon Dioxide Analyzer the overall depth of th...

Страница 45: ...hown above is an explosion proof oxygen measuring system The Series 9510 Carbon Dioxide Analyzer is identical to that shown above in terms of mechanical layout APPENDIX D WIRING INSTRUCTIONS FOR EXPLO...

Страница 46: ...APPENDIX E ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 40...

Страница 47: ...WALL MOUNTED ELECTRONICS WITH REMOTE SENSOR ENCLOSURE ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 41...

Страница 48: ...G3 Alpha Omega Instruments Corp IN2 10915 9510 Remote WIRING TO REMOTE SENSOR ENCLOSURE 11076 A REMOTE SENSOR V INTERFACE BOARD LOCAL ENCLOSURE REMOTE SENSOR SERIES 9510 REMOTE ENCLOSURE 0 PULSE 4 20...

Страница 49: ...mote enclosure shown is identical to the one used for the Series 9510 Carbon Dioxide Analyzer with one exception The Series 9510 enclosure will not have the Isolation valves as shown in the above draw...

Страница 50: ...APPENDIX H EXPLOSION PROOF REMOTE SENSOR ENCLOSURE SHOWING INTERNALLY MOUNTED ELECTRICAL TERMINAL BLOCK ALPHA OMEGA INSTRUMENTS CORP SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL 44...

Отзывы: