Wall Mounted Electronics with Remote Sensor Enclosure.
Please refer to Appendix D
3.1.3 Initial Check
The Series 9510 Carbon Dioxide Analyzer is ready to use out of the shipping container. The analyzer has
been calibrated at the factory and, under normal conditions, re -calibration is not required at initial start-up.
In the case of explosion proof systems, some wiring to meet local electrical codes will be required Please
refer to Section 4.0 for details regarding routine calibration checks.
Optional Battery Operation
NOTE FOR BATTERY OPERATION
If the Series 9510 has been equipped with optional NiMH batteries, it is designed to provide up to eight
hours of operation on a fully charged set of batteries (without pump). With a pump, one may expect to
obtain up to 3-4 hours of continuous operation before charging of the batteries will be necessary. The
batteries should be fully charged upon receipt of the analyzer. If not, the user should recharge the batteries
upon initial startup by powering the analyzer from AC power. Allow at least 1 6 hours of recharging time.
NOTE: THE POWER ON SWITCH MUST BE KEPT IN THE “ON” POSITION DURING CHARGING.
The battery option should never be used when the analyzer is powered from a separate DC source. This will
cause the batteries to overcharge which may cause the batteries to explode. Be certain all wiring (alarms,
remote sensor, RS-232C, etc.) has been done before applying AC power to the instrument
ALPHA OMEGA INSTRUMENTS CORP.
SERIES 9510 CARBON DIOXIDE ANALYZER INSTRUCTION MANUAL
12