Page 18 of 33
500101
Read L -
80.0
Lower Limit
Read
H
20.0
Higher Limit
ACAL
10.0
AutoCal Value
SPC
Pt
20.0
Single Point Calibration Action Point.
SPC
Va
20.0
Single Point Calibration Offset Value.
Units
ºC
Currently Set Engineering Units
P
Unit
barg
Currently Set Pressure Units
Press
00.0
Currently Set Pressure Value in Current Units
Sensor
B
Type of Sensor fitted to SADPminiEx
4mA -
80.0
Low mA output in Currently Set Engineering Units.
20mA
20.0
High mA output in Currently Set Engineering Units.
Epot
384
Position of the electronic AutoCal potentiometer
Exit
Exit option
11.1.10
Temperature
Description – This function allows the user to see the current temperature within the
SADPminiEx head in both ºC and ºF.
Operation – In the ‘Setup Menu’ select the ‘Temperature’ option.
11.1.11
Exit
Description – The Exit option allows the user to return to the normal display.
Operation – In the ‘Setup Menu’ select the ‘Exit’ option.
11.2
The Logging Menu
(Full logging versions only)
NOTE: This internal function can be performed in hazardous areas.
Transferring data to a PC cannot be performed in hazardous areas - See section 17, page 25.
See also Appendix C for SADPminiEx Logging Menu Map.
The Logging menu it activated by pressing the ‘
’ and ‘
’ keys simultaneously. From
this menu, the user has the choice of several options detailed below: