Product Manual - MED Multi-well Connector (MED-C11) 1
1. MED Multi-well Connector Components and Functions
1. MED Multi-well Connector Components and Functions
The MED Multi-well Connector (MED-C11) is the Connector for the MED64-Allegro System. It connects
the MED Multi-well Probe to the MED64 Head Amplifier (MED-A64HE1S) allowing recording extracellular
signals from 4 (or 8) cultured samples in a MED Multi-well Probe.
Figure 1.
System diagram for the MED64-Allegro System.
1-1. Components and their functions
MED Multi-well Connector unit
Figure 2.
MED Multi-well Connector. Front view (left) and back view (right).
(1) Ground terminal
Connect to the GND terminal on the MED64 Head Amplifier with the green grounding wire.
(2) Screw hole
Use these hole to install the Connector onto the MED64 ThermoBase.
(3) Probe port
Install the MED Multi-well Probe here.
(4) Output terminal
Connect to the INPUT 1-16CH, 17-32CH, 33-48CH, 49-64CH on the MED64 Head Amplifier with the
MED Connector Cable (20 pins).
MED Multi-well MED Multi-well MED64 Head Amp. MED64 Main Amp. PC and Mobius
Stimulator
Digitizer
Probe Connector
(1) (2) (3) (4)