6
Sample Entry
To calibrate the moisture, enter the moisture values in the left side of the display. After
the proper moisture value has been entered, press the right arrow key and move the
brackets around the 0000 value. Place the grain from the holding hopper into the weigh
bucket containing the probe.
MST 1 New 1
12.5 [ 0000 ]
Press the right arrow again and the raw data will appear on the right side of the display.
The value is not a real-time reading, so the right arrow button needs to be pressed again
to verify a stable value being displayed.
MST 1 New 1
12.5 [ 0548 ]
Press the EXE key to accept the stable raw data value. After the value has been
accepted, the SS-LRX will be advance to the next moisture sample. The ‘MST 1’
indicates the same data curve is being calibrated and the ‘New 2’ represents the second
point of the calibration.
MST 1 New 2
[ 0.00] 0000
Repeat this procedure with all of the grain samples. The accuracy of the curve will be
dependent upon the number of samples and the accuracy of the known sample moisture.
When the calibration process is complete, press ESC key three times to return to the
Main Menu.
Manual Entry
To calibrate the moisture with known values, enter the moisture on the left side of the
display. After the proper moisture value has been entered, press the right arrow key and
move the brackets around the 0000 value.
MST 1 New 1
12.5 [ 0000 ]
Using the keypad, manually enter the known raw moisture reading.
MST 1 New 1
12.5 [ 0548 ]
Содержание Seed Spector LRX
Страница 2: ...SEED SPECTOR LRX OPERATORS MANUAL...
Страница 4: ......
Страница 36: ...28...
Страница 37: ...SEED SPECTOR LRX MANUAL 29...
Страница 38: ...30...
Страница 39: ...SEED SPECTOR LRX MANUAL 31...
Страница 40: ...32...
Страница 41: ...SEED SPECTOR LRX MANUAL 33...
Страница 42: ...34...
Страница 43: ...SEED SPECTOR LRX MANUAL 35...
Страница 44: ...36...
Страница 45: ...SEED SPECTOR LRX MANUAL 37...
Страница 46: ...38...
Страница 47: ...SEED SPECTOR LRX MANUAL 39...