              SEED SPECTOR LRX MANUAL 
 

 

5

 
With the arrow highlighting MST 1, press the EXE key to display the calibration menu. 
 

!

!

!

!

 Initialize 

     Generate 

 
A new moisture calibration requires the user to select and execute ‘

Initialize

’ after which 

the data type had been selected.  The Initialize feature erases the old calibration from the 
SS-LRX memory.  If the initialization is not done, the new moisture points will be added to 
the existing calibration causing inaccuracies.  Highlight the Initialize option and press the 
EXE key. 
 

!

!

!

!

 Init MST 1 ? 

     NO 

 
Press the down arrow key to move the selection arrow to ‘

YES’

 and press the EXE key.  

After ‘MST 1’ has been initialized, the SS-LRX returns to the calibration menu screen. 
 

     NO 

!

!

!

!

 

YES 

 
The SS-LRX is ready to be calibrated with the known samples.  At the calibration menu, 
highlight the ‘

Generate

’ and press the EXE key to display the calibration entry screen.  

The ‘MST 1’ indicates the data being calibrated and the ‘New 1’ represents the point of 
the calibration. 
 

MST  1      New  1 
[       0.00]     0000 

 
To calibrate the moisture reading the user will need to use several samples of grain, each 
with a known moisture percentage.  The moisture contents of these samples should be 
determined using the ASAE dry down method or a calibrated moisture meter such as the 
Dickey-john GAC 2000.  To develop the most accurate curve inside the Seed Spector 
LRX, it’s recommended that the lowest moisture sample be inputted first.  The next 
sample needs the highest moisture, which should represent the highest moistures 
measured in the field during harvest.  After these two samples have been entered, the 
remaining grain samples can be input at random.  A minimum of two samples is required 
for a valid calibration. Three or four samples shoud be used, if a wide range of moistures 
is being harvested. Adjacent samples should be no closer than 3-4% in absolute 
moisture. 

Содержание Seed Spector LRX

Страница 1: ...Operator s Manual for Seed Spector LRX 99 M Avenue Nevada IA 50201 1549 Phone 515 382 3506 Fax 515 382 2973 www almaco com MOE0108022...

Страница 2: ...SEED SPECTOR LRX OPERATORS MANUAL...

Страница 3: ...signs in good condition Replace missing or damaged safety signs Be sure new equipment components and repair parts include the current safety signs SAFETY CAUTION Read the following guidelines careful...

Страница 4: ......

Страница 5: ...g with in the field versatility The result of our efforts and years of development with practicing agronomists has been a laboratory instrument capable of volume production yet still meeting the rigid...

Страница 6: ...3 DATA TYPE MENU 3 CALIBRATION MENU 3 POWER UP 3 CALIBRATION PROCEDURE STANDARD INTERFACES 3 MT1 MT2 3 WT1 WT2 4 TM1 TM2 4 TWT1 TWT2 4 MOISTURE CALIBRATION 4 SAMPLE ENTRY 6 MANUAL ENTRY 6 WEIGHT CALIB...

Страница 7: ...ermined as if you are sitting in the operator s seat Record Keeping is an important part of machine operation This information will help when ordering replacement parts and during service calls Please...

Страница 8: ...ed on ALMACO products such accessories being subject to the warranties of their respective manufacturer ALMACO shall in no event be liable for consequential damages or contingent liabilities arising o...

Страница 9: ...for a dual plot application are displayed on the Seed Spector screen The top row of data is the information for the left plot and the bottom for the right plot A typical application for this selection...

Страница 10: ...displayed in red on the keypad The ALT key is not operational in this version of the LRX software ON Turns the power on the Seed Spector LRX To turn the LRX off the ALT and EXE keys need to be pressed...

Страница 11: ...the power switch The following screens will appear sequentially System Initializing Seed Spector LRX by ALMACO Version 1 25 25 Copyright 2002 Moisture Weight 4 5 0 00 CALIBRATION PROCEDURES The calib...

Страница 12: ...and moisture chamber from the holding hopper This is the same as if the combine were harvesting plots 3 The combine should be running the same as during harvest That is the shakers header cylinder op...

Страница 13: ...EXE key to display the calibration entry screen The MST 1 indicates the data being calibrated and the New 1 represents the point of the calibration MST 1 New 1 0 00 0000 To calibrate the moisture read...

Страница 14: ...the value has been accepted the SS LRX will be advance to the next moisture sample The MST 1 indicates the same data curve is being calibrated and the New 2 represents the second point of the calibra...

Страница 15: ...readings The rods should be in the horizontal position To check for friction press down on the weigh hopper and release With the weight displayed on the Seed Spector observe the stabilized weight read...

Страница 16: ...alibrated with the known samples At the calibration menu highlight the Generate and press the EXE key to display the calibration entry screen The WGT 1 indicates the data being calibrated and the New...

Страница 17: ...the right side of the display The value is not a real time reading so the right arrow button needs to be pressed again to verify a stable value being displayed WGT 1 New 2 10 02 2987 Press the EXE ke...

Страница 18: ...N INITIALIZATION OF THE RAM WILL DELETE ALL CALIBRATION FILES FROM THE SEED SPECTOR LRX To initialize the RAM Disconnect power from the Seed Spector Remove the cover remove the 8 screws that retain th...

Страница 19: ...8 K2com Relay 2 COM 9 K3no Relay 3 N O 10 K3nc Relay 3 N C 11 K3com Relay 3 COM 12 K4no Relay 4 N O 13 K4nc Relay 4 N C 14 K4com Relay 4 COM 15 NO CONNECTION 16 NO CONNECTION J2 REMOTE PORT CONNECTOR...

Страница 20: ...IELD 8 10 TO 5 VDC J4 WEIGHT PORT CONNECTOR Left or Right SOURCE PIN CONNECTION DESCRIPTION 1 EXE 2 EXE 3 SIG 4 SIG 5 SHIELD J5 EXPANSION PORT CONNECTOR requires optional relay board SOURCE PIN CONNEC...

Страница 21: ...SEED SPECTOR LRX MANUAL 13 J7 PRINTER PORT CONNECTOR 9 PIN FEMALE DSUB SOURCE PIN CONNECTION DESCRIPTION 2 PPRXD 3 PPTXD 5 SG Signal Ground...

Страница 22: ...grain Check sensor connector for moisture Printer fails to print Check that printer is turned on Examine printer for paper jam loose ribbon etc Check baud rate setting Check cable connections at print...

Страница 23: ...SEED SPECTOR LRX MANUAL 15 Bucket holds in open position and will not close Bad connector Bad park switch in wiper motor Bucket door jammed by grain or debris...

Страница 24: ...SET GATE THRESHOLD PARAMETER FOR ANALOG CHANNELS 20 H SET GATE THRESHOLD COUNT PARAMETER FOR ANALOG CHANNELS 5 I INITIALIZE DEVICE CALIBRATION J SET CYCLE COMMAND STRING CCS K L M DEVICE GAIN N DEVIC...

Страница 25: ...LTER Syntax A D N Where D is the device channel 0 7 N is the averaging value 0 99 Default S W initialization values will be 5 This command sets the AVERAGING Filter parameter for the specified Device...

Страница 26: ...0 1200 2400 4800 9600 19 2K 38 4K 57 6K 115 2K 230 4K Response current value new value CR LF C DOWNLOAD CALIBRATION FILE Syntax C D Where D is the device channel 0 7 This command informs the SS LRX th...

Страница 27: ...e of the current A D value that Device data filtering is active If the absolute difference between 2 consecutive A D values exceeds the G value then filtering is set to not active Note See Legend for...

Страница 28: ...ce AD is the previous A D reading for the Device G is the averaging threshold value N is the averaging threshold count Note See Legend for Device D location mapping Response H D current value new valu...

Страница 29: ...W300Z1 NOTE The is not included into string in the SS LRX Port Command S W Response No response when a string parameter is sent A J with no string parameter requests the current J string S CR LF M SET...

Страница 30: ...This command sets the TARE parameter for the specified Device The Tare Parameter is the value used by the SS LRX to calculate the displayed value For a weight channel this value is typically the empt...

Страница 31: ...2 4 1 3 MST 2 4 2 4 TEMP 1 3 3 5 TEMP 2 4 3 6 TWT 1 5 1 7 TWT 2 5 2 8 Prompt ASCII 36 Where S is the command string Response No response when a string parameter is sent A Q with no string parameter r...

Страница 32: ...LY 3 b2 b2 3 RLY 4 b3 b3 4 RLY 5 NA b4 5 RLY 6 NA b5 6 RLY 7 NA b6 7 RLY 8 NA b7 8 RLY 9 NA b8 9 RLY 10 NA b9 To direct the command to the IRB vs the XRB DIP switch 7 value 64 will be assigned to indi...

Страница 33: ...tored into theAnlaog array for Device D TV is the Previous Tare Value in theAnalog array for Device D CV is a Current Value for Device D T D Example CV 1 2 TV 34 7 V Default 0 TV CV Default TV TV 1 2...

Страница 34: ...or Device D location mapping Response RV CR LF W WAIT Syntax W V Where V is a value representing hundredths of a second This command causes the SS LRX to wait before executing the next command in a st...

Страница 35: ...This command sets the SCAN time for a specific channel This is the time base used by the SS LRX to read the A D converter Response Y D current value new value CR LF Z RESET RELAY Syntax Z N Where N is...

Страница 36: ...28...

Страница 37: ...SEED SPECTOR LRX MANUAL 29...

Страница 38: ...30...

Страница 39: ...SEED SPECTOR LRX MANUAL 31...

Страница 40: ...32...

Страница 41: ...SEED SPECTOR LRX MANUAL 33...

Страница 42: ...34...

Страница 43: ...SEED SPECTOR LRX MANUAL 35...

Страница 44: ...36...

Страница 45: ...SEED SPECTOR LRX MANUAL 37...

Страница 46: ...38...

Страница 47: ...SEED SPECTOR LRX MANUAL 39...

Страница 48: ...ALMACO has Service 99 M Avenue Nevada Iowa 50201 www almaco com Phone 515 382 3506 Fax 515 382 2973 Manufacturers of Specialized Agricultural Equipment...

Отзывы: