![Allwin CS2676FX Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/allwin/cs2676fx/cs2676fx_user-manual_2904422014.webp)
CS2676FX Insulation Resistance Tester User Manual
Ver:1.1
Page 14 of 25
3.1
illustration of front panel
1
、
POWER SWITCH
Press the button to turn on the power, pop-up for turning off the power
.
2、
STOP key
It is used to stop the test during the test; In additions, it is used to stop alarming when the
DUT is failed to test.
3、
START key
In the
“stop” status, press the key to start to test
.
4
、
Gear select key
(1)
Voltage select
Select the test voltage 500V/1000V according to the actual requirements.
(2)
Range switch
After the test voltage is selected, the corresponding voltage indicator lights, if you are
not sure about the resistance range, please select the maximum range 2000M, and then
select the appropriate range according to the display resistance, If the resistance is greater
than the selected range, the screen displays "----", please switch to the high range to
measure; If the resistance is less than the selected range, the tester will do the failure
alarm
.
5、
Fail indicator
If the resistance value is less than the range, the indicator will be light and the beep will be
alarmed
.
6、
Test indicator
During the test, the indicator will be light, the high voltage port output the high voltage. Please do
not touch the high voltage output terminal.
7、Screen
Display the resistance value of the DUT.
8、
500V
voltage indicator
9、
1000V
voltage indicator