19
3.3 Preparing the measurement
Select the smallest slot suitable for the test sample diameter and rotate the fixture so the selected slot is in
the 3 o’clock position, closest to the wire clamp fixture.
Place the cable connection into the adapter, so that the sleeve of the cable connector stays securely inside
the ring and cannot be pulled out through the slot.
The wire clamp fixture should be positioned in the full open position.
The manual pull tester needs a free length of the cable off minimum 14cm (5.5“), the motorized version a
length of 10cm (4“).
Soft cable coatings may slide through the clamp fixture or be pulled out of the crimp sleeve. In this case
please cut off the coating in 4cm to 8cm (2...3“) distance to the cable joint under test.