9.0 Trouble shooting (FAQ) 

Does the instrument work also after a 
being stored in the dark for a longer 
period? 

Yes. Even after a self discharging of the internal energy storage the instrument can be used immediately. A back-
up battery saves guaranties the basic function of the device; see chap. 2.3. 

The instrument shows already minute 
values before the cable is fixed! 

The electronic captures minute changes after taring and those are shown in the PEAK display (<1,5N). This does 
not affect the accuracy of measurement as the PEAK register is refreshed always with the highest captured value 
while pulling off the terminal. 

The peak value cannot be saved because 
the display shows 0,0 after the break! 

You may have changed the mode from PEAK measurement with dragging function to steady update of the actual 
value (PEAK symbol disapears). Delete the memory by pressing the 0-Button and change the mode back to PEAK 
with dragging function by pressing the S-Button until the PEAK symbol shows up. 

The instrument shows no consistent 
values! 

Please pay attention to torn parts of wire in the spaces around the connection adapter; should this occur, blow 
them off or remove them using the vacuum cleaner.. 

The display shows “OvErL“! 

The  load  exceeds  the maximum range. Release the load and inspect the instrument. If no consistent value is 
shown it might be necessary to replace the load cell. 

The USB connection is interrupted! 

Please check whether FMT-W_Connect incl. the USB driver is installed and the USB-ports can be used. 

The display does not light up and the drive 
does not run. 

Please check power connection. In case of drive failure (e.g. overheating) the red LED beside the DC socket will be 
illuminated. Switch the instrument off, and wait appr. 15min until the drive cooled down and all capacitors 
discharged. If the LED still lights up when switching on the instrument please contact our technical service. 

A loud vibrating sound occurs but the 
clamping device does not turn and the pull 
force does not increase. 

They drive might be blocked and overloaded! Switch the instrument off and remove the blockage. Restart the 
instrument, the clamping device will return to its home position. 

The measuring does not start immediatly 
after pressing the I-button. 

The drive performs a reference run (very slow speed) after switching the instrument on or after a measuring cycle. 

The buzzer has no sound! 

Please check whether the buzzer is activated (submenu P17=1) and a limit is setted. 

The clamping device returns to its home 
pos, although the joint did not break or 
the limit is not reached. 

The real time force slipped under 25% of the setted maximal force limit (e.g. the cable coating slides or peels off) 
or the limit switch for the final position of the clamping device was reached. 

The cable slides through the clamping 
device or the coat peels off! 

Remove the cable coating in 4 to 8cm distance to the cable joint. (See also chapter 3.3). 

 

10.0 Accessories

 (Exchangable terminal fixtures) 

 

 

 

 

The scope of supply comprises a standard terminal adapter, premounted in our works. If your testing material requires special adapters for the 
terminal of your cable, you may find one of the options below useful. In order to change the rotating adapter easily you place a 4mm pin or M4 
screw (max. 25mm (1”)) into the hole and rotated it to a 3 o’clock position. Now the rotation is blocked and you are able to open the screw in the 
center. Replace the terminal adapter and tighten the screw before your release the pin or M4 screw. 

 

 

 
Pin for ring terminals 
Diameter > 3,5mm 
Part.No.: FMT-931 

 

Roller grip 1kN, Opening 0…7mm 
for pull test of welded wire and cable connections 
 
Part.No.: FMT932WT 

 

Slotted terminal adapter 
For mini-crimps with beveled contour 
Part.No.: FMT-953 

 

 
L-Crimp-adapter 
For 3,5mm L-crimps (other sizes on request) 
 
Part.No.: FMT-932 

 

Raw body for terminal adapters 
Adapter for own machining of special 
applications 
Part.No.: FMT-951 

 

 
Adapter for ring terminals. Turntable with 8 pins 
(D=2|3|4|5|6|8|10|12mm). 
 
Part.No.: FMT-956 

 

Содержание FMT-W30C5

Страница 1: ...re Terminals Dynamom tre Numerique par Jonctions de C bles Dinamom trico Digital Por cables con unions Adecuadas Dinamometri Digitales a di Cavi con Connettori BDA Version 2 4 Bedienungsanleitung Oper...

Страница 2: ...l tester FMT W30 20 3 6 Start measuring AutoTara with motorized pull tester FMT W40 21 3 7 Automatic memorizing of measured values with motorized pull tester FMT W40 21 3 8 Display statistical data 22...

Страница 3: ...formation 2 1 Scope of Supply unpacking and Setting Up The scope of supply consists of Measuring instrument with integrated electronics Hand lever or roller grip USB Interface cable for optional softw...

Страница 4: ...ecommended date of next calibration and the nominal measuring range Fn To turn the instrument off you may press the I Button for 2 seconds else the AUTO OFF function will turn the instrument off autom...

Страница 5: ...he cable connector stays securely inside the ring and cannot be pulled out through the slot The wire clamp fixture should be positioned in the full open position The manual pull tester needs a free le...

Страница 6: ...e develops which works upon the cable connection The peak value which mostly appears before the connection snaps up is captured at 1000Hz and is indicated in the display After 5 seconds of unchanged m...

Страница 7: ...tive if 25 of the force limit minimum tensile strength P3 or a force of 40N is reached If the actual measured force falls below this level the cable clamp returns to its home position and the measurin...

Страница 8: ...etting or load new test limits The memory and static function can be turned off see chap 6 0 2S 4 Tensile force limit control Depending on the wire strength and the related standard the following tens...

Страница 9: ...not apply with the motorized version FMT W40 The lower limit is adjusted automatically to 25 or max 40N of the upper limit With a short press on the S Button you return to the main setting menu I 0 Pr...

Страница 10: ...icense file has to be generated and saved in the Alluris folder Please enter the received license key and the serial no of your pull tester and click generate license file The license key has to match...

Страница 11: ...standard requiring 30 N the instrument is parametrized to 36 N This allows pulling further while testing than the limit set to compensate for a relaxing of the test object The test report refers to t...

Страница 12: ...mpany details with your own logo etc The template is filed in C Alluris FMT W_Connect Formular xlsx Fields with gray background are filled by the data entered in FMT W_Connect and the data of the test...

Страница 13: ...to the main setting menu 0 I 1 0 The pull tester tares and resets the PEAK automatically with start of each measurement in order to eliminate any influences by change of the material and temperature I...

Страница 14: ...a PC 24 VDC 3A universal power supply Standby consumption 42mA Average consumption drive active 0 8A Interface USB 2 0 Optional software FMT W_Connect Temperature range Operation 5 40 C 5 30 C Storage...

Страница 15: ...29 8 2 Register for updates In order to receive information about product changes and updates and to enjoy full warranty please register your product at www alluris de service php...

Страница 16: ...service A loud vibrating sound occurs but the clamping device does not turn and the pull force does not increase They drive might be blocked and overloaded Switch the instrument off and remove the blo...

Страница 17: ...irm additionally that the product complies with the requirements of the EMC Directive 2006 42 EEC In bereinstimmung mit der WEEE Richtlinie 2002 96 EC ist dieses Ger t eingestuft als Monitoring and Co...

Страница 18: ...nischen Bundesanstalt PTB Braunschweig im Rahmen des DAkkS zur ckf hren A 2 Calibration Confirmation acc DIN EN 10204 2 1 We hereby confirm in accordance to DIN EN 10204 2 1 that this instrument has b...

Страница 19: ...75 Alluris GmbH Co KG Basler Strasse 63 65 DE 79100 Freiburg Germany info alluris de www alluris de Fon 49 0 761 47979 3 Fax 49 0 761 47979 44...

Страница 20: ...65 DE 79100 Freiburg Germany info alluris de www alluris de Fon 49 0 761 47979 3 Fax 49 0 761 47979 44 BDA_FMT W Art Nr 50337 Stand 07 15 Alluris GmbH Co KG Alle Rechte vorbehalten Alluris ist eine ei...

Отзывы: