16
General settings
To customise the instrument to the application, some of the factory settings (defaults) can be changed. These
adjustments will be saved in the microprocessor and be recalled when you switch on the force gauge the next
time.
Switching off the temperature compensation (Default: SET)
The temperature drift is compensated automatically. In general, it is recommended to have the this function
turned on as long as it does not effect your measuring results. Only when measuring very minute forces over a
longer period of time at a slow rate and under stable environmental conditions, it might be useful to switch the
tracking off.
•
Switch off the instrument.
•
Press and hold PEAK- and MODE-buttons.
•
Press ON/OFF-button.
•
Wait until TRK oFF is displayed.
•
Release PEAK- and MODE-buttons.
Selecting the baud rate of the RS232C interface (Default: 2400)
The speed of the data transfer for the RS232C interface can be adjusted to match your data acquisition. This is
possible by changing the baud rate:
•
Switch off the instrument.
•
Press and hold RESET-button.
•
Press ON/OFF-button.
•
Wait until f01 is displayed, then release RESET-button.
•
Select function f04 by repeating to press the PEAK-button until f04 is shown on the display.
•
Select appropriate baud rate (2400/4800/9600/19200) by repeating to press the MODE-button.
•
Press RESET-button to save your settings.
Switching off Auto Power-Off function (Default: 10 min)
Under the battery operation the instrument will be automatically switched off after approx. 10 minutes. This
Auto Power-Off function can be deactivated, which may become necessary when monitoring the fluctuation of
applied forces for a longer period of time. This function is not active, when the instrument is powered by the
AC-adaptor.
•
Switch off the instrument.
•
Press and hold RESET-button.
•
Press ON/OFF-button.
•
Wait until f01 is displayed, then release RESET-button.
•
Select function f03 by repeating to press the PEAK-button until f03 is shown on the display.
•
Select setting (10/OFF) by repeating to press the MODE-button.
•
Press RESET-button to save your settings.
Change the refresh rate of the display (Default: 3/sec)
The force applied to the load cell of the instrument is internally processed at a rate of 1000Hz, which allows the
precise capturing of the peak values. The readings on the display, however, are refreshed 3 times/sec only for
the convenience of the human eyes. You may increase or decrease the factory settings for them to match your
personal demands.
•
Switch off the instrument.
•
Press and hold RESET-button.
•
Press ON/OFF-button.
•
Wait until f01 is displayed, then release RESET-button.
•
Select function f02 by repeating to press the PEAK-button until f02 is shown on the display.
•
Select refresh-time (1/2/3/5/10/20 times/sec) by repeating to press the MODE-button.
•
Press RESET-button to save your settings.
TRK
oFF
Содержание FMT-100WT
Страница 2: ...2 ...
Страница 3: ...3 Deutsch 4 English 14 Francais 24 Espanol 34 Italiano 44 ...
Страница 55: ...55 ...
Страница 56: ...56 ...