Alltion (Guangxi) Instrument Co., Ltd.

AM-2000

User’s Manual | Safety Caution

3

Storage Environment

Ambient temperature range: -40

℃~

+55

Relative humidity range: 10% ~ 60% RH

Atmospheric pressure range: 500hPa ~ 1060hPa

Well ventilated indoor environment free of corrosive gas or other harmful substances.

Safety Characteristics

Adaptor: Input: DC 12V / 3A

Classification according to the degree of protection of the liquid: IPX0;

Standard IEC60601-1 Type II Equipment.

Safety Requirements on Installation and Use

Safety Requirements

This instrument can be used only for the purposes described in the User's Manual.

Only trained and instructed personnel are allowed to use this instrument. The customer
or the organization operating the equipment has the responsibility to train and guide all
personnel using the equipment.

Before starting the instrument, please completely comprehend the User's Manual,
including the User's Manual for accessories and other system components.

Keep the User's Manual in order to facilitate the operator to read at any time.

Please observe all symbols and labels of the instrument!

The modification and repair of this instrument can only be performed by the ALLTION
service personnel or others authorized by ALLTION.

Do not place any container filled with liquid over the instrument. Confirm that no liquid
can penetrate into the instrument.

No modification of this equipment unless authorized by the manufacturer.

Although the equipment conforms to the intent of the standard IEC 60601-1-2 in relation to
electromagnetic compatibility, electrical equipment may produce interference. If interference
is suspected, move equipment away from sensitive device or contact us.
Do not store or use the instrument in a damp room. Do not expose the instrument in the
place with splashing, dropping or water mist.
When the instrument is generating smoke, electric spark or a strange noise, please
immediately cut off power supply of the instrument. Do not use this instrument until it has
been repaired by our service agent.
Please note that local regulations take precedence over the requirements of the above
mentioned criterions. If you have any enquiry, please contact the local ALLTION dealer.

Requirements on Installation

The installation of the product will be completed by our service representative or by
professional personnel authorized by us. Please make sure that the following operational
requirements have been met:

All the safety-related mechanical connections (please find the Manual for details) are
properly connected and all screws have been tightened.

Содержание AM-2000

Страница 1: ...AM 2000 Surgical Microscope User s Manual Please read the instruction carefully before using it PA1805 UM01 EN Version A4 2019 12 25 Alltion Guangxi Instrument Co Ltd...

Страница 2: ...k Wuzhou Guangxi China E mail sales alltion com or sales alltion microscope com Tel 86 774 2836101 Fax 86 774 2836192 Postal Code 543100 Website http www alltion com After Sale Service Provider Alltio...

Страница 3: ...of Fixed Floor Stand 13 Installation of Ceiling Mount 14 Installation of Low Position Wall Mount 15 High Position Wall Mount 16 Installation of Cross Arm System 17 Upright Installation of Cross Arm 17...

Страница 4: ...s declaration electromagnetic emission 37 Guidance and manufacturer s declaration electromagnetic immunity for all EQUIPMENT and SYSTEMS 38 Guidance and Statement from Manufacturer Electromagnetic Imm...

Страница 5: ...user of the instrument or providing the useful information to the user Information Symbols The item symbols used in this User s Manual are defined as follows The content of the upper and lower items i...

Страница 6: ...ed bracket optional cross arm bracket including electric installation and microscope head including visual optical component microscope body objective lens and illuminating system Range of Application...

Страница 7: ...e personnel or others authorized by ALLTION Do not place any container filled with liquid over the instrument Confirm that no liquid can penetrate into the instrument No modification of this equipment...

Страница 8: ...ient Do not disassemble or assemble the supporting components during use to prevent the Support system from being out of balance damaging the components or hurting the patient This surgical microscope...

Страница 9: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Safety Caution 5 1 4 3 2 3 5...

Страница 10: ...straight or 45 bevel 1 3 Eyepiece 12 5X or 10X 2 4 Magnification pod five range zooming 1 5 Control handle 1 6 Objective lens F250 objective lens optional F300 objective lens please find the Optional...

Страница 11: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Product Assembly 7 2 1 3 4 5 7 8 6...

Страница 12: ...e binocular head Straight tube 2 45 binocular head 45 inclined 3 200 300 Variable Objective Lens Continuous zoom from 200mm to 300mm 4 F200 objective lens F200 5 F250 objective lens F250 6 F300 object...

Страница 13: ...evice 13 Camera adapter Interfaces with Sony Canon Nikon cameras optional 14 2 D Rotation Binocular Assistant Scope Connector 15 3 D Rotation Binocular Assistant Scope Connector 16 ALL CAM2 Full Funct...

Страница 14: ...Alltion Guangxi Instrument Co Ltd 10 Product Assembly AM 2000 User s Manual 17 19 20 18...

Страница 15: ...not provided if any component is not provided please contact the local dealer in time Please check the product if there s any damage especially the optical components if any please contact the local d...

Страница 16: ...put it aside Put base 1 on the horizontal ground and lock four casters Fit the plastic cover 2 onto the base 1 and tighten the compression ring 3 Insert the upright post 4 in the installation hole of...

Страница 17: ...d with 14 driller the hole depth is 75mm and the hole position is as shown on the right diagram 170 170 14 If the position and dimension of the installation hole are incorrect the stand can t be insta...

Страница 18: ...ram When drilling please note the direction of power cord connection The power connector is along same direction with the pre installed socket on the ceiling Lock the ceiling mount assembly 2 to the c...

Страница 19: ...t shall be smooth enough otherwise it will cause the product to be tilted after installation Drill 6 holes in the wall with 14 driller the hole depth is 75mm The position and dimension of the hole is...

Страница 20: ...shall be smooth enough otherwise it will cause the product to be tilted after installation Drill 6 holes in the wall with 14 driller the hole depth is 75mm The position and dimension of the hole is as...

Страница 21: ...ing bolts 2 on the cross arm remove the plastic cover 1 and put it aside Lift the cross arm over the pole cross the power cable 3 from the mounting hole of cross arm 4 as shown by the bellow arrow Ins...

Страница 22: ...the column insert the decorative ring 2 into the column insert the power plug 3 on the column into the power connector 4 Tighten with the nut 5 on the power plug Insert the connecting shaft on the fi...

Страница 23: ...ctive lens 300mm objective lens 350mm objective lens 400mm objective lens and 200 300mm objective lens are the same Mount the large objective 1 into the microscope mount with two M4 screws 3 Install t...

Страница 24: ...ion Align the binocular tube 1 at the positioning pin and install it in the bayonet of microscope body 3 Tighten the locking screw 2 Before loosening the grip please make sure the binocular head barre...

Страница 25: ...Product Assembly 21 Wiring Connect the attached power wire according to the following picture If the specification of power cord plug does not match the local socket standard please contact your loca...

Страница 26: ...ob of each joint is valid judge whether the locking function is normal Turn on the power switch observe whether the indicator of the power switch turns on whether the light is illuminated from the mic...

Страница 27: ...on method is the same as the Installation of Binocular Head Barrel see Page 20 Installation of Objective Lens For objective lens of other specifications the installation methods are the same as Large...

Страница 28: ...ctions Product Components 1 Main body of microscope 2 180 binocular and eyepiece 3 Objective lens 4 120 hanger bracket 5 Cross arm bracket 6 Mobile floor stand 7 Fixed Stand Mount 8 Low position wall...

Страница 29: ...interface For 12V 1 0A DC output 3 Lighting shift knob For switching different illumination means large spot without color filter means medium spot without color filter means small spot without color...

Страница 30: ...provide diopter compensation at 6D and 6D Setting the diopter adjustment at 0D if the operator wear glasses Rotating the diopter adjustment to the best position till you see the most clear view if the...

Страница 31: ...Alltion Guangxi Instrument Co Ltd AM 2000 User s Manual Product Functions 27 Objective lens 1 Focusing knob It is used for manually setting the image definition focusing working distance 1...

Страница 32: ...rotation of the main lens of the microscope as indicated by the arrow on the left 2 Friction adjustment knob for tilting motion left right of the main lens of the microscope Used to adjust the fricti...

Страница 33: ...ll accessories adjust the balance of the balance arm with the knob the adjusting identification as the one in the left picture For rotating the knob easily the balanced arm shall be in the horizontal...

Страница 34: ...Alltion Guangxi Instrument Co Ltd 30 Product Functions AM 2000 User s Manual 1 2 3 4 5 6 7 7 7...

Страница 35: ...sories adjust the balance of the balance arm with the knob the adjusting identification as the one in the left picture For rotating the knob easily the balanced arm shall be in the horizontal position...

Страница 36: ...Alltion Guangxi Instrument Co Ltd 32 Product Functions AM 2000 User s Manual 5 7 3 1 2 4 6 7...

Страница 37: ...cope Check before use There must be no patient when checking the microscope To terminate the operation of the equipment please turn off the power switch or disconnect from the network power supply Che...

Страница 38: ...ched Adjust the surgical microscope to the maximum magnification and move the surgical microscope to the position that has clearest image Clearer image can be obtained by fine adjustment of the focus...

Страница 39: ...Instrument Co Ltd AM 2000 User s Manual Performance Parameters 35 Performance Parameters Basic dimension The basic dimensions of the whole machine and the motion range of the joint are shown in the f...

Страница 40: ...shift Orange filter green filter large spot without filter medium spot without filter small spot without filter Optical parameters Large objective lens F200 F250 F300 F350 F400 Variable magnific ation...

Страница 41: ...decreasingly in electromagnetic compatibility transmission and immunity performance to use or replace with the accessories with non matched design Do not replace the components without authorization...

Страница 42: ...V differential mode 1 kV differential mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply in...

Страница 43: ...smitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in metres m b Field strengths from fixed RF transmitters as determined by an electromagnetic s...

Страница 44: ...ration distance according to frequency of transmitterm 150kHz 80MHz d 3 5 V1 p 80MHz 800MHz d 3 5 E1 p 800MHz 2 5GHz d 7 E1 p 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12...

Страница 45: ...t free box Do not use any chemical cleaning agent corrosive solvent or detergent with scratching effect which will damage the surface of the optical device The stain blood stains etc on the surface of...

Страница 46: ...ntermittently during use The cooling window and the air inlet are covered or blocked by external object Remove the foreign object and clean the cooling window Failure of cooling fan Contact the local...

Отзывы: