Routine Maintenance
Particle Filter Replacement
•
The MCV100(B) contains a particle filter located inside the air inlet on the
side of the unit. This filter cleans the ambient air drawn in by the
compressors and also cleans the air used in the function of the blender.
This filter should be checked every 4 months and changed if visibly dirty.
To replace the filter, remove the protective screen by prying off the cap
with a fingernail or sharp object. Once the screen is removed, the foam
filter can be removed and replaced.
34
Protective Screen
(covers foam particle filter)