Initial Operation and Check-Out
Completion and Troubleshooting
•
Before powering off the vent, return the unit to the following settings
–
BPM = 10
–
Tv = 600ml
–
High Airway Pressure Alarm Setting ~ 40 cm H
2
O
–
Low Airway Pressure Alarm Setting ~ 9 cm H
2
O
–
Adjustable Airway Pressure Relief = 60 cm H
2
O
–
Source Gas Selection = 100% air
–
Turn the power off by pressing and holding the power button for one second
•
This routine procedure should ensure that the ventilator is suitable for use
each time it is deployed.
•
Should the unit fail any of the tests, contact the Allied Healthcare Products,
Inc. Technical Service Department at 800-411-5136 for assistance.
32