Camera Link registers
56
Goldeye Camera Link Register Controls Reference V1.0.0
RegDeviceBaudRateSwitchConfirmTimeout
Timeout for a confirmation write while switching the data transmission rate to a
new value. Applies to GenCP link on Camera Link devices. Default is 250
milliseconds according to GenCP standard. When set to 0xFFFFFFFF, the
confirmation write requirement is disabled.
RegDeviceManufacturerInfo
Part code and flags of the camera.
Example entry:
Goldeye G-008|4068080|
RegDeviceModelName
Camera family and model name, such as “Goldeye CL-032“. Software should use
the
DevicePartNumber
to distinguish between models.
Register name
RegDeviceBaudRateSwitchConfirm
Timeout
Register type
IntReg
Access mode
R/W
Unit
Milliseconds
Address
0x0000B800
Length (Bytes)
4
Bits used
31..0
Possible values
Description
250
Default
Register name
RegDeviceManufacturerInfo
Register type
StringReg
Access mode
R
Address
0x00000104
Length (Bytes)
64
Register name
RegDeviceModelName
Register type
StringReg
Access mode
R
Address
0x00000044
Length (Bytes)
64