GUI reference
R&S
®
ZND
356
User Manual 1173.9557.02 ─ 61
Coupled Markers
. The label indicates the selected
:
"Coupling Type"
Label
"All"
"Coupled Markers"
"Channel"
"Coupled Markers in <name of active channel>"
"Diagram"
"Coupled Markers in <name of active diagram>"
Coupling also works if
Marker <m>
in
one
trace. The
Marker
<m>
of the other traces then follow the movements of the tracked marker. The same
holds true for the reference marker.
Remote command:
CALCulate:MARKer:COUPled[:STATe]
Coupling Type
Determines the set of traces whose markers are coupled to the markers of the active
trace when
is enabled.
●
"All"
: marker coupling applies to all traces in the active recall set that have the
same stimulus variable as the active trace
●
"Channel"
: marker coupling applies to all traces in the active channel
●
"Diagram"
: marker coupling applies to all traces in the active diagram that have
the same stimulus variable as the active trace
The selected "Coupling Type" applies to all marker coupling relations in the current
recall set. Changing the "Coupling Type" automatically disables marker coupling
throughout the recall set.
Remote command:
Reset Marker Coupling
Convenience function for disabling
"All" (default).
Remote command:
CALCulate:MARKer:COUPled[:STATe] OFF
CALCulate:MARKer:COUPled:TYPE ALL
Marker softtool
www.allice.de
Allice Messtechnik GmbH