Command reference
R&S
®
ZND
1041
User Manual 1173.9557.02 ─ 61
Parameters:
<Slope>
Port-specific slope factor
Range:
-40 dB/GHz to +40 dB/GHz. The resulting power
range over the entire frequency sweep must be
within the power range of the analyzer; refer to the
data sheet.
Increment: 0.1-dB/GHz
*RST:
0 dB/GHz
Default unit: DB/GHZ
Example:
See
SOURce<Ch>:POWer<PhyPt>[:LEVel][:
Manual operation:
See
SOURce:POWer:REDuce:SDELay
<SettlingDelay>
If power reduction at sweep end is enabled (
ON) this defines the settling delay at subsequent sweep starts, i.e.
the time between power-up and sweep start when a new sweep is requested.
This is a global setting.
Parameters:
<SettlingDelay>
Settling delay
Default unit: s
Example:
See
SOURce<Ch>:POWer<PhyPt>:REDuce[:STATe]
<ReduceAtSweepEnd>
Deprecated
; superseded by
Enables/disables power reduction at sweep end.
This is a global setting.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the setting is valid for all
channels.
<PhyPt>
Physical port number. This suffix is ignored; the setting is valid
for all ports.
Parameters:
<ReduceAt
SweepEnd>
OFF
(default): at sweep end the output power of the sweep start
is restored
ON
: at sweep end the output power of the driving port is reduced
as if the channel base power was set to its minimum possible
value
SCPI command reference