Command reference
R&S
®
ZND
1043
User Manual 1173.9557.02 ─ 61
Note:
If the start power entered is greater than the current stop power
(
SOURce<Ch>:POWer<PhyPt>:STOP
), the stop power is set to the start power plus
the minimum power span (increment) of 0.01 dB.
If the stop power entered is smaller than the current start power
(
SOURce<Ch>:POWer<PhyPt>:STARt
), the start power is set to the stop power
minus the minimum power span of 0.01 dB.
SOURce:POWer:SWEepend:MODE
<SweepEndModes>
Selects the power mode at sweep end.
This is a global setting.
Parameters:
<SweepEndModes> AUTO | REDuce | KEEP
AUTO: at sweep end restore the power at sweep start
REDuce: at sweep end reduce the output power as if the chan-
nel base power was set to its minimum possible value
KEEP: at sweep end keep the power at its current level
Example:
SOURce:POWer:SWEepend:MODE REDuce
Reduce power at sweep end.
SOURce:POWer:SWEepend:SDELay 0.01
Use a settling delay of 10 ms
SOURce:POWer:SWEepend:MODE AUTO
At sweep end, restore the power at sweep start
Manual operation:
SOURce:POWer:SWEepend:SDELay
<SettlingDelay>
If sweep end mode REDuce or KEEP is active (see
) this defines the settling delay at subsequent sweep starts, i.e. the time between
power reset and sweep start when a new sweep is requested.
This is a global setting.
Parameters:
<SettlingDelay>
Settling delay
Default unit: s
Manual operation:
"Settling Delay / Reset Delay"
7.3.15.7
SOURce... (other)
SOURce<Ch>:LPORt<LogPt>
<PhysicalPort>[, <PhysicalPort>]
Assigns the logical port number <LogPt> either to the single-ended physical port
<PhysicalPort1> or to a pair of physical ports <PhysicalPort1>,<PhysicalPort2>, at the
same time defining them as balanced port.
SCPI command reference