The performance of the Test Strips has been evaluated in clinical tests.
Accuracy:
The user performance for the Blood Glucose Monitoring System was evaluated
by using capillary blood samples from 405 patients with those obtained using a YSI Model
2300 STAT Plus Glucose Analyzer.
For glucose concentration < 75mg/dL
Site
Within ± 5 mg/dL
Within ± 10 mg/dL
Within ± 15 mg/dL
1/1 (100%)
7/7 (100%)
7/7 (100%)
15/15 (100%)
1/1 (100%)
7/7 (100%)
7/7 (100%)
15/15 (100%)
For glucose concentration ≥ 75mg/dL
Site 1
Site 2
Site 3
Total (3 Site)
0/1 (0%)
7/7 (100%)
5/7 (71.4%)
12/15 (80%)
Site
Site 1
Site 2
Site 3
Total (3 Site)
Within ± 5 %
103/142 (72.5%)
118/161 (73.3%)
71/87 (81.6%)
292/390 (74.9%)
Within ± 10 %
137/142 (96.5%)
156/161 (96.9%)
87/87 (100%)
380/390 (97.4%)
Within ± 15 %
142/142 (100%)
161/161 (100%)
87/87 (100%)
390/390 (100%)
Within ± 20 %
142/142 (100%)
161/161 (100%)
87/87 (100%)
390/390 (100%)
47
Performance Characteristics