16
Applying blood and reading results
Apply the round drop of blood to the side edge of the yellow window of the
test strip and then blood drop will be absorbed into the narrow channel.
Touch and hold the blood drop to the side edge of the yellow window of test
strip until blood completely fills the confirmation window and the meter beeps.
Then, your meter will start the countdown.
Apply the Blood Sample
You should apply blood to the left side edge of the test strip
as directed.
- Do not put blood on the top of the test strip.
- If insufficient blood is inserted, error message (Er8)
will be displayed.
As you can see in the left illustration, you have to insert
a sufficient amount blood in the blood checking area.
- Even if the meter begins the countdown despite the
fact that the confirmation window is not completely
filled, do not apply more blood to the test strip.
Discard the test strip and repeat the test with a new
test strip.
- If blood is inserted in the opposite direction, error
message (Er9) will be displayed.
Er8
Er9
Performing the Blood Glucose Test with
Incorrect
Correct
Correct