Remote control commands
R&S
®
NGU
116
User Manual 1179.2531.02 ─ 03
.
Parameters for setting and query:
<arg0>
IN | EXT
IN
Pin 3 of DIO connector is monitored.
EXT
Pin 2 (Ch1) of DIO connector is monitored.
Example:
TRIG:SOUR:DIO:PIN IN
Pin 3 of DIO connector is monitored for trigger source "Digital In
Channel".
TRIGger[:SEQuence][:IMMediate]:SOURce:OMODe
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce:OMODe?
<arg0>
Sets or queries the operation mode to trigger on for trigger source "operation mode"
.
Parameters for setting and query:
<arg0>
CC | CV | CR | SINK | PROTection
CC
If respective channel operation mode is detected in CC mode,
corresponding trigger-out parameters are triggered.
CV
If respective channel operation mode is detected in CV mode,
corresponding trigger-out parameters are triggered.
CR
If respective channel operation mode is detected in CR mode,
corresponding trigger-out parameters are triggered.
SINK
If respective channel operation mode is detected in sink mode,
corresponding trigger-out parameters are triggered.
PROTection
If respective channel operation mode is detected in protection
mode (OVP, OCP, OPP OTP), corresponding trigger-out param-
eters are triggered.
Example:
TRIG:SOUR OMOD
TRIG:SOUR:OMOD:CHAN OUT1
TRIG:SOUR:OMOD CV
If CV mode is detected on Ch1, a trigger event is activated to
corresponding trigger-out parameters.
Trigger commands
www.allice.de
Allice Messtechnik GmbH