Remote control commands
R&S
®
NGU
115
User Manual 1179.2531.02 ─ 03
7.4
Trigger commands
The
TRIGger
subsystem contains the commands for signal triggering.
TRIGger[:SEQuence][:IMMediate]:SOURce
...................................................................... 115
TRIGger[:SEQuence][:IMMediate]:SOURce:DIO:PIN
......................................................... 115
TRIGger[:SEQuence][:IMMediate]:SOURce:OMODe
......................................................... 116
TRIGger[:STATe]
<arg0>
Enables or disables the trigger system.
Upon being triggered, the selected trigger source
on page 115 becomes active.
See
Setting parameters:
<arg0>
1
Enables the trigger system.
0
Disables the trigger system.
TRIGger[:SEQuence][:IMMediate]:SOURce
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce?
<arg0>
Sets or queries the trigger source.
See
Parameters for setting and query:
<arg0>
OUTPut | OMODe | DIO
OUTPut
Trigger source is from the output channel.
OMODe
Trigger source is from the different modes (CC, CR, CV, Sink,
OVP, OCP, OPP and OTP) detected from the output channel.
DIO
Trigger source is from DIO connector at the instrument rear
panel.
Example:
TRIG:SOUR OMOD
Trigger source "Operation Modes" is selected for monitoring.
TRIGger[:SEQuence][:IMMediate]:SOURce:DIO:PIN
<arg0>
TRIGger[:SEQuence][:IMMediate]:SOURce:DIO:PIN?
<arg0>
Sets or queries the DIO pin to trigger on for trigger source "Digital In Channel".