7
Installation
Plumbing
All L Series Meters and LC Series Controllers are equipped with female inlet and outlet port connections.
Because the flow meters set up a laminar flow condition within the flow body, no straight runs of pipe
are required upstream or downstream of the meter. The inlet and outlet ports are equal in size and
symmetric (in-line). The port sizes (process connections) and mechanical dimensions for different flow
ranges are shown on page 29.
Meters/controllers with M5 (10-32) ports have o-ring face seals and require no further sealant or tape.
On other meters/controllers, avoid the use of pipe dopes or sealants on the ports as these compounds
can cause permanent damage to the meter should they get into the flow stream. Use of thread sealing
Teflon tape is recommended to prevent leakage around the threads. When applying the tape, avoid
wrapping the first thread or two to minimize the possibility of getting a piece of shredded tape into the
flow stream. When changing fittings, always clean any tape or debris from the port threads.
It is also recommended that a 20 micron filter be installed upstream of meters/controllers with full scale
ranges of 100CCM or less and a 50 micron filter be installed upstream of meters/controllers with full
scale ranges above 100CCM.
Note:
Avoiding long runs of small diameter tubing upstream or downstream of controllers will
reduce water hammer.
Mounting
All L Series Meters and LC Series Controllers have mounting holes for convenient mounting to flat
panels. These meters are position insensitive and can be mounted in any orientation. The sizes and
dimensions for the mounting holes are shown on pages 30-33.
Note:
For applications that may continuously introduce occasional air bubbles to the flow stream
upstream of the device, the device may be mounted upside down to prevent the bubbles from becoming
trapped in the differential pressure sensor ports.
If the device has been installed upside down, avoid
using the bleed screws as water may leak into the electronics housing causing permanent damage that
is not covered under warranty!
Application
Maximum operating line pressure is 100 PSIG (689 kPa).
Caution: Exceeding the maximum specified
line pressure may cause permanent damage to the solid-state differential pressure transducer.
If the line pressure is higher than the 100 PSIG (689 kPa), a pressure regulator should be used upstream
from the flow meter to reduce the pressure to 100 PSIG (689 kPa) or less if possible. Although the
meter’s operation is uni-directional, reversing the flow direction will inflict no damage as long as the
maximum specified limits are not exceeded. The differential pressure sensor utilized in this flow meter/
controller is a very sensitive device capable of detecting minute differences in pressure.
Avoid installations (such as snap acting solenoid valves upstream) that apply instantaneous
high pressure to the meter as permanent damage to the differential pressure sensor could
result. This damage is not covered under warranty!