Maintenance and Recalibration
General:
L and LC Series Water Flow Meters and Controllers require minimal maintenance. The primary
cause of damage and/or long-term inaccuracy in these devices is contamination and/or corrosion
damage. Water should be filtered for particulates or biological materials that may grow in the device.
When removing these units from the line for any extended period of time, make an effort to remove all
of the water from the device, as deposits of calcium or other soluble minerals can affect the accuracy
of the device.
Recalibration: The recommended period for recalibration is once every year. Providing that the CLEAN,
DRY, and NON-CORROSIVE mantra is observed, this periodic recalibration is sufficient. A label located
on the back of the meter lists the recalibration due date. The meter should be returned to the factory for
recalibration near the listed due date. Before calling to schedule a recalibration, please note the serial
number on the back of the meter. The Serial Number, Model Number, and Date of Manufacture are also
available on the Manufacture Data 2 screen (page 16).
Cleaning: L and LC Series Flow Meters require no periodic cleaning. If necessary, the outside of the
meter can be cleaned with a soft dry rag. Avoid excess moisture or solvents.
For repairs, recalibrations, or recycling of this product, contact:
Alicat Scientific, Inc.
2045 N Forbes Blvd. Bldg.103
Tucson, Arizona 85745
USA
Ph. 520-290-6060
Fax 520-290-0109
email: [email protected]
Web site: www.alicatscientific.com
CCM
CCH
LPM
LPH
CFM
CFH
CFH
0.0021
0.00003
2.1189
0.035
60.0
1.0
CFM
0.000035 0.0000005 0.035
0.00059
1.0
0.0166
LPH
0.06
0.001
60.0
1.0
1699.0
28.316
LPM
0.001
0.000017
1.0
0.0166
28.316
0.4719
CCH
60.0
1.0
60000.0
1000.0
1699011.0 28317.0
CCM
1.0
0.0167
1000.0
16.667
28317.0
471.947
Flow Conversion Table: