2 Safety
All warnings in the manual are summarised in this section.
Pay special attention to the instructions below so that severe personal injury and/or damage to the pump are avoided.
“b1” control system requirements
Measurement system shall be chosen according to EN-ISO 80079-37.
Always test and validate the control system before production start-up.
The ignition protection system performance requirements:
Temperatures:
-
+/-2°C
-
30s reaction time
Flow:
-
+/- 10%
-
30s reaction time
The demand for the ignition protection system b1 must be according to ISO 13849-1 PL c cat. 2 or IEC 61508 SIL 1 and
must be calibrated and functionally tested on a yearly basis.