INSTRUCTION MANUAL
p. 12/13
ALERT
TM
J5
p. 13/13
MAINTENANCE & TROUBLESHOOTING
Troubleshooting Table
IF THE ALERT J5
THEN
Does not power on
Change the batteries
Displays
Not Activated
Press the left button
Cannot print serially (using
communication cable)
• Check that the printer is on
• Check the cable connection
• Check the cable
Is displaying irregular
results
• Change the mouthpiece. Use
part number
95-000130
, to
order more mouthpieces from
ACS
• Eating, drinking or smoking 20
minutes prior to taking the test
can contaminate results. Wait
for at least 20 minutes and
repeat the test
• Check if a calibration is due
Has a clogged or dirty
mouthpiece
• Change the mouthpiece
• Use part number
95-000130
to order more mouthpieces
from ACS
Is using a printer that is
out of paper
• Replace the paper roll
• Use part number
95-000171
to order more rolls from ACS
Is displaying
PEAK ERROR
• You are trying to calibrate the
device with a solution less
than 100 BAC
• Do not attempt to calibrate
yourself
Requires service
Contact ACS for return procedure