Maintenance Instructions

Routine Testing

It is recommended to test the system weekly by simulating spasm movements
on  the  bed  for  15  seconds  (the  default  time  unless  changed  by  a  user  via
programming) & ensure the alarm call is sent to the pager.

How to Clean the Devices

It is also recommended to regularly clean both units by wiping with cotton wool
pads moistened (compressed until dripping stops) with a mild detergent (0.5%
washing up liquid) solution or by using an alcohol or baby wipe. Avoid getting
any liquid into the micro-USB connector.

Changing Battery’s in the Companion

mini

Remove the base of the monitor as per figure 3 above OR figure 7 & 8 below.
Also please notice the polarity marked into the case [Fig. 5] when replacing the
2x AAA batteries.

After changing the battery’s, the monitor MUST be turned on again by pressing
the recessed TEST button and waiting for the 60 second test mode to finish.

Figure 5.

Figure 6.

Page | 11

Содержание Companion Mini

Страница 1: ...Wireless Epilepsy Support mini...

Страница 2: ...on of seizures and the safeguard of individuals Please read this quick start guide carefully to ensure the correct set up use and maintenance of your new product for years to come If you have any ques...

Страница 3: ...the Pager 6 Positioning and Setup 7 System Operation 8 Alarm Mode 8 Pager Charging 10 Pairing Instructions 10 Pairing the Monitor to the Pager 10 Maintenance Instructions 11 Routine Testing 11 How to...

Страница 4: ...ypes Tonic clonic Partial Generalised The carer is then alerted via the Alert iT Safelink Radio System This offers a failsafe detection of radio integrity using a Class 2 radio receiver The carer must...

Страница 5: ...Pager 1x P176 Companion Mini Monitor 2x Spare AAA LR03 Batteries 1x P153 Pager Power Supply 2x Cable ties Velcro Pads 1x USB Stick containing Quick start video programming software Instructions 1x USB...

Страница 6: ...is woken up and placed into a test mode by pressing the recessed test button on the front of the monitor Figure 1 typically using a pen tip The internal red light will flash to indicate the unit is n...

Страница 7: ...nd all three LEDs will flash The green Pager LED will blip every 8 seconds to show that the pager is awake and listening for alarm signals If the monitor is turned on you now see the red Monitor LED o...

Страница 8: ...Companion Mini detects movement If you are happy that the monitor is picking up each press tap of the mattress then you can secure it into positon using the supplied Velcro pads or cable ties To use t...

Страница 9: ...le using the provided software contact Alert it for support Alarm Mode When prolonged movements are detected the Alarm LED on the pager will light up it will make an audible alarm and will vibrate Pre...

Страница 10: ...ff No Three Beeps 3 Flashes 3 Flashes 3 Flashes Switching On No Once Flash Once Flash Once Flash Once Pager battery getting low No No Flash every Second N A N A Pager battery getting very low No Sound...

Страница 11: ...he Pager Ensure the monitor is turned on Pairing is only possible within 30 seconds of powering up the pager To pair the Monitor to the Page place the monitor close to the pager and press and hold the...

Страница 12: ...compressed until dripping stops with a mild detergent 0 5 washing up liquid solution or by using an alcohol or baby wipe Avoid getting any liquid into the micro USB connector Changing Battery s in th...

Страница 13: ...line for the purpos es of fault diagnosis and the provision of advice Repair or if this is not possible replacement of faulty equipment free of charge however the equipment must be returned to Alert i...

Страница 14: ...x AAA LR03 Non rechargeable Battery Life 12 months Operating Temp 5 35 C Humidity 90 max Non condensing Sealing IP53 not waterproof The system complies with 93 42 EEC as a Class 1 Medical Device for u...

Страница 15: ...onduct a risk assessment to determine if the level of reliability offered by the monitor is sufficient or if additional monitoring is needed Contact the manufac turer for assistance with Risk Evaluati...